The effect of natural rapid cooling and oven slow cooling on the precision of thermoluminescence measurements of LiF:Mg,Ti is investigated. Three separate series of measurements resulted in average precisions of 5.1 and 5.0%, respectively. However, the highest precision of 1.7% (1 SD) was achieved for an oven-cooled material.