Aims: The Naviswiss system (Naviswiss AG, Brugg, Switzerland) is a handheld imageless navigation device used to improve the accuracy of implant positioning in total hip arthroplasty (THA). However, clinical data for leg length discrepancy and femoral offset is lacking, and the validity of the system has not been reported for patients undergoing THA in the lateral decubitus position. This study aimed to report the accuracy of the device in this patient population.

Methods: Patients underwent THA in the lateral decubitus position performed by a single surgeon. Component position measured by the device intraoperatively was compared to postoperative measurements on computed tomography (CT) scans. Agreement between the navigation system and postoperative measurements was reported for acetabular cup inclination, acetabular cup version, femoral offset, and leg length discrepancy.

Results: Thirty-three patients were included in the analysis. The mean difference between intraoperative and postoperative CT measurements was within 2° for angular measurements and 2 mm for leg length. Absolute differences in the two indices were up to 4° and 3 mm. The mean bias was 1°-2° overestimation for cup orientation and up to 2 mm overestimation for leg length change. However, 95% limits of agreement did not exceed absolute thresholds of 10° and 10 mm, especially after correction for bias. One case (3%) was declared intraoperatively for issues with fixation on the greater trochanter.

Conclusions: The accuracy of the Naviswiss system falls within clinically acceptable recommendations for acetabular cup placement, femoral offset, and leg length for total hip arthroplasty with a anterolateral approach in lateral decubitus position. The system could be further improved with regression-based bias correction.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10773062PMC
http://dx.doi.org/10.1186/s42836-023-00224-0DOI Listing

Publication Analysis

Top Keywords

leg length
20
lateral decubitus
16
decubitus position
16
total hip
12
hip arthroplasty
12
femoral offset
12
postoperative measurements
12
acetabular cup
12
imageless navigation
8
navigation system
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!