Comparison of Point Accuracy Between Two Widely Used Continuous Glucose Monitoring Systems.

J Diabetes Sci Technol

Texas Diabetes and Endocrinology, Round Rock, TX, USA.

Published: May 2024

Background: Safe and effective self-management of glucose levels requires immediate access to accurate data. We assessed the point accuracy of the Dexcom G7 Continuous Glucose Monitoring System (Dexcom, Inc., San Diego, CA, USA) and FreeStyle Libre 3 (Abbott Diabetes Care, Alameda, CA, USA) sensors in a head-to-head comparison.

Method: Multicenter, single-arm, prospective, nonsignificant risk evaluation enrolled adults (≥ 18 years) with diagnosed type 1 diabetes (T1D) or type 2 diabetes (T2D). Accuracy was assessed by comparing sensor data to laboratory reference values Yellow Springs Instrument [YSI] and capillary blood glucose values. Outcome measures were differences in mean absolute relative difference (MARD), number and percentage of matched glucose pairs within ±20 mg/dL/±20 of reference values within glucose ranges: < 54, 54 to 69, 70 to 180, 181 to 250, > 250 mg/dL, and combined.

Results: Data from 55 adults were included in the analysis. Analysis showed significantly lower MARD with the FreeStyle Libre 3 sensor vs the Dexcom G7 sensor (8.9% vs 13.6%, respectively, < .0001) with a higher percentage of glucose values within ±20 mg/dL/±20 of reference (91.4% vs 78.6%). The MARD values for both continuous glucose monitoring (CGM) sensors were similar during the first 12 hours; however, the FreeStyle Libre 3 MARD was notably lower than the Dexcom G7 MARD during the next 12 hours (10.0% vs 15.1%, respectively, < .0001) and throughout the study period.

Conclusions: The FreeStyle Libre 3 sensor was more accurate than the Dexcom G7 sensor in all metrics evaluated throughout the study period. This is the first head-to-head study to our knowledge that compares the flagship products currently in widespread use of the two largest CGM manufacturers.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC11089878PMC
http://dx.doi.org/10.1177/19322968231225676DOI Listing

Publication Analysis

Top Keywords

freestyle libre
16
continuous glucose
12
glucose monitoring
12
point accuracy
8
glucose
8
type diabetes
8
reference values
8
glucose values
8
±20 mg/dl/±20
8
mg/dl/±20 reference
8

Similar Publications

Accuracy and feasibility of continuous glucose monitoring system in pancreatectomy patients.

Langenbecks Arch Surg

January 2025

Department of Surgery, Division of HBP Surgery & Liver Transplantation, Korea University College of Medicine, Seoul, Korea.

Purpose: Pancreatectomy patients often experience challenging fluctuations in blood glucose levels; therefore, they require a reliable monitoring system. This study aimed to determine the accuracy and acceptability of a continuous glucose monitoring (CGM) system compared with the intermittent capillary glucose test in patients who have undergone pancreatectomy.

Methods: Thirty non-diabetic pancreatectomy patients participated.

View Article and Find Full Text PDF

The FreeStyle Libre (Abbott) is a human continuous glucose monitor sensor that has been used to monitor glucose concentrations in dogs and cats with diabetes. This sensor and other similar continuous glucose monitoring sensors are increasingly popular in veterinary medicine, specifically for management of diabetes mellitus. The sensor decreases veterinary visits and reduces the stress (and expense) of multiple blood sample collections needed to produce glucose curves.

View Article and Find Full Text PDF

Introduction: Glycated haemoglobin (HbA1c) is currently the gold standard for assessing glycaemic control in diabetes, given the established relationship with microvascular and macrovascular complications in this condition. However, HbA1c is affected by non-glycaemic factors, while also failing to provide data on hypoglycaemic exposure and glucose variability, which are associated with adverse vascular outcomes. Continuous glucose monitoring (CGM)-derived glucose metrics provide a more comprehensive assessment of glycaemia, but their role in predicting future vascular complications remains unclear.

View Article and Find Full Text PDF

Background: Continuous glucose monitors (CGMs) are used to characterize postprandial glucose responses and provide personalized dietary advice to minimize glucose excursions. The efficacy of such advice depends on reliable glucose responses.

Objectives: To explore within-subject variability of CGM responses to duplicate presented meals in an inpatient setting.

View Article and Find Full Text PDF

Continuous glucose monitoring (CGM) has emerged as a superior method to glycated hemoglobin (HbA1c) monitoring for glycemic control assessment in type 1 diabetes (T1D). The association between CGM parameters and diabetic kidney disease (DKD) has not been extensively researched. The aim of the present study was to compare CGM metrics between patients with stable and progressive DKD and T1D.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!