Background: The aim of the study was to answer the question of whether flattening filter (FF) and flattening filter-free (FFF) beams can be used alternately in the volumetric modulated arc therapy (VMAT) treatment technique, regardless of the size of the irradiated volume [small (S) or large (L) planning target volume (PTV)].
Materials And Methods: Two groups of patients were examined: a group with a S-PTV-laryngeal cancer and a group with a L-PTV - gynecological volume. For each patient, two treatment plans were made for beams (energies): FFF-X10MV and FF-X15MV. Then, a statistical analysis, nonparametric test, and independent groups were performed, comparing the beams' impact on the analyzed treatment plans.
Results: In the case of laryngeal irradiation (S-PTV), there are no statistically significant differences between the energy used and the assessed parameters of the plan. In the case of gynecological volume (L-PTV), only statistically significant differences were noted for the number of monitor units depending on the energy used. For a large irradiated volume (gynecological case), the use of FFF beams increases the number of monitor units by 39,4% in relation to the FF beam.
Conclusions: In the case of gynecological neoplasms, statistically significant differences were found in the number of monitor units. Therefore, in the case of irradiation of L-PTV, it is recommended that flattening-filtering beams are used due to the smaller number of monitors. In the case of S-PTV, no statistically significant differences were found between the types of beams used (FF or FFF) and the treatment plan parameters analyzed in the study.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10764046 | PMC |
http://dx.doi.org/10.5603/rpor.97508 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!