Background/objective: During the initial stages of rehabilitation after anterior cruciate ligament (ACL) surgery, a pivotal role is played in ensuring effective recuperation and averting complications. An often-employed strategy to tackle ACL laxity during this period involves the incorporation of synthetic materials for reinforcement. The objective of this study is to compare the effectiveness of conventional suture tape and multiple high-strength sutures as augmentation techniques for ACL repair.

Methods: Ten preserved cadaveric knees embalmed using the Thiel method were segregated into two groups, each containing five knees. In one group, traditional suture tape was employed for augmentation, while the other group utilized multiple high-strength sutures. Each knee underwent a cyclic load of 1000 sine wave cycles, succeeded by an axial distraction load until failure ensued. The resultant displacement and ultimate load at failure were assessed to contrast the efficacy of the two augmentation techniques.

Results: The group utilizing multiple high-strength sutures exhibited a significantly higher load to failure at time-zero (1690.7 N) compared to the suture tape group (987.6 N) (P = .003). Furthermore, the multiple high-strength sutures group demonstrated significantly reduced displacement after 1000 cyclic loads (6.6 mm) in comparison to the suture tape group (16.3 mm) (P < .001).

Conclusions: Multiple high-strength sutures show better biomechanical properties for the augmentation of ACL repair at time-zero. Both suture tape and multiple high-strength sutures had ultimate load-to-failure values higher than the natural ACL loads. Therefore, these substances might serve as augmentation options to prevent the ACL's gradual elongation, a critical concern particularly in the initial stages of rehabilitation.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10755459PMC
http://dx.doi.org/10.1016/j.asmart.2023.11.007DOI Listing

Publication Analysis

Top Keywords

suture tape
20
multiple high-strength
20
high-strength sutures
20
load failure
12
traditional suture
8
sutures augmentation
8
anterior cruciate
8
cruciate ligament
8
tape group
8
group
6

Similar Publications

Unlabelled: Chronic lateral ankle instability is a common diagnosis in foot and ankle clinics. Internal Brace (IB) augmentation is a surgical procedure that utilizes fibertape augmentation of the lateral ankle ligaments. Studies have shown the superiority of fibertape augmentation over traditional lateral ankle stabilization procedures such as the Brostrom or Brostrom-Gould.

View Article and Find Full Text PDF

Background And Aims: Interest in repairing ruptured anterior cruciate ligaments (ACL) has resurged, with suture tape augmentation (internal brace, ACL-IB) emerging as a prominent technique. However, the efficacy of ACL-IB compared to ACL reconstruction (ACL-R) remains unclear. We conducted a meta-analysis to address this gap.

View Article and Find Full Text PDF

The posterior cruciate ligament (PCL) is the strongest ligament of the knee and plays an important role in stabilizing the knee joint posteriorly. PCL tears are common injuries in sports injuries and traffic accidents; however, clinical outcomes after PCL reconstruction have not yet met clinicians' expectations, with a high postsurgery failure rate reported. Suture tape augmentation and supplementary fixation have shown ideal biomechanical properties in early studies.

View Article and Find Full Text PDF

This article aims to present a comprehensive technical note detailing our preferred treatment approach for tibial tuberosity avulsion fractures in the adult and elderly populations, particularly in scenarios characterized by low tissue quality and limited bone stock. Existing literature on this fracture type is scarce, with many described techniques relying on optimal bone quality for effective screw fixation of the tibial tuberosity. Various methods for tibial tuberosity avulsion fixation include K-wires, cannulated screws, staples, tension bands, suture anchors, and in select cases, direct transosseous sutures.

View Article and Find Full Text PDF

Purpose: The purpose of this study was to evaluate the biomechanical properties of SutureTape as an alternative technique for arthrodesis of the metacarpophalangeal (MCP) and proximal interphalangeal (PIP) joint arthrodesis when compared with surgical steel wire.

Methods: A total of 32 fingers (index, long, ring, and small) from two matched pair cadaveric hands were used. K-wire and surgical steel wire were used for MCP and PIP joint arthrodesis of the control group (group I), whereas K-wire and SutureTape were used for the experimental group (group II).

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!