Background: There are limited data on the impact of intracardiac echocardiography (ICE)-guided transcatheter aortic valve replacement (TAVR) on the new permanent pacemaker implantation (PPMI) rate.

Objectives: This study investigated the feasibility and outcome of transjugular ICE (TJ-ICE) -guided TAVR, by visualizing the relationship between the membranous septum (MS) and the transcatheter aortic valve (TAV).

Methods: Among patients with severe aortic stenosis who underwent TAVR between February 2017 and June 2020, this study enrolled a total of 163 patients with TJ-ICE-guided TAVR. MS length was measured by ICE. The primary endpoint of this study was the incidence of new PPMI at 30 days.

Results: The mean age of the patients in this study was 84.9 ± 4.6 years, and 71.2% of the patients were female. Device success was 96.3% with TJ-ICE guidance. A TJ-ICE-related complication occurred in 1 case (0.6%). The median length of the MS was 5.8 mm (IQR: 5.0-6.9 mm). Excellent intraobserver (intraclass correlation coefficient [ICC]: 0.94; 95% CI:0.79-0.98;  < 0.001) and interobserver (ICC: 0.93; 95% CI: -0.05 to 0.98;  < 0.001) agreements were shown. The new PPMI rate was 6.7% at 30 days without a significant difference between balloon-expandable valves and self-expandable valves (3.4% vs 8.7%;  = 0.226). Patients with a TAV implantation depth less than MS length had a significantly lower incidence of new PPMI compared with patients with a TAV implantation depth greater than MS length (2.1% vs 13.4%;  = 0.005), regardless of baseline right bundle branch block presence (6.7% vs 66.7%;  = 0.004) or absence (1.2% vs 8.2%;  = 0.041).

Conclusions: TJ-ICE-guided TAVR demonstrated remarkable feasibility and safety. The TJ-ICE-guided final TAV position had a significant impact on the new PPMI rate. (Tokai Valve Registry; UMIN000036671).

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10751646PMC
http://dx.doi.org/10.1016/j.jacasi.2023.07.013DOI Listing

Publication Analysis

Top Keywords

transcatheter aortic
12
aortic valve
12
feasibility outcome
8
outcome transjugular
8
valve replacement
8
tj-ice-guided tavr
8
incidence ppmi
8
 < 0001
8
ppmi rate
8
patients tav
8

Similar Publications

Re-sheathing failure with Navitor during transcatheter aortic valve implantation: a case report.

Eur Heart J Case Rep

January 2025

The Second Department of Internal Medicine, University of Toyama, 2630 Sugitani, Toyama, Toyama 930-0194, Japan.

Background: Self-expanding valves used in transcatheter aortic valve implantation (TAVI) are designed to allow recapture and repositioning, facilitating optimal placement and mitigating conduction disturbances and paravalvular leakage. Here, we present a rare case in which the Navitor (Abbott Structural Heart, Santa Clara, CA, USA) could not be recaptured.

Case Summary: An 81-year-old Japanese woman with very severe aortic stenosis and a massively calcified nodule at the non-coronary cusp (NCC) underwent TAVI with a 25 mm Navitor valve.

View Article and Find Full Text PDF

Background: Since the transcatheter valve-in-valve (ViV) procedure was introduced in 2007, a few cases of infective endocarditis (IE) following the ViV procedure have been reported, which can be predisposed by older age, pre-existing medical conditions, and procedural techniques. Paravalvular abscesses constitute a rare complication of IE, resulting from extending IE beyond the valve annulus, less commonly caused by species. This complication is more common in prosthetic valves, particularly bioprosthetic valves.

View Article and Find Full Text PDF

Blunt thoracic aortic injury treated with thoracic endovascular aortic repair in hybrid emergency room: A case report.

Trauma Case Rep

February 2025

Department of Emergency and Critical Care Medicine, Hyogo Emergency Medical Center, 1-3-1 Wakinohamakaigandori, Chuo-ku, Kobe, Hyogo 651-0073, Japan.

Background: Hybrid emergency rooms (ERs) allow computed tomography (CT) scanning, interventional radiology, and surgery all in the same suite. Severe trauma patients with blunt thoracic aortic injury (BTAI) require rapid diagnosis and treatment. Hybrid ERs allow the potential for clinicians to implement multiple therapeutic procedures, including thoracic endovascular aortic repair (TEVAR), for these types of conditions without the need to transport the patients.