Fully automated dual-run manufacturing of [C]PIB on FASTlab.

Nucl Med Biol

Vanderbilt University Institute for Imaging Science, Vanderbilt University Medical Center, Nashville, TN, USA; Center for Molecular Probes, Vanderbilt University Medical Center, Nashville, TN, USA; Department of Radiology and Radiological Sciences, Vanderbilt University Medical Center, Nashville, TN, USA. Electronic address:

Published: February 2024

This report describes an updated, fully automated method for the production of [C]PIB on a cassette-based automated synthesis module. The method allows for two separate productions of [C]PIB, both of which meet all specification for use in clinical studies. The GE FASTlab developer system was used to create the cassette design as well as the controlling tracer package. The method takes 16 min from the delivery of [C]MeOTf to the FASTlab, or 35 min from the End of Bombardment; and reliably produces 3547 ± 586 MBq of [C]PIB in high radiochemical purity (> 98 %). This methodology increases the production capacity of radiopharmaceutical facilities for [C]PIB, and can easily produce 4 batches in a single day with limited infrastructure footprint.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10922476PMC
http://dx.doi.org/10.1016/j.nucmedbio.2023.108873DOI Listing

Publication Analysis

Top Keywords

fully automated
8
[c]pib
5
automated dual-run
4
dual-run manufacturing
4
manufacturing [c]pib
4
[c]pib fastlab
4
fastlab report
4
report describes
4
describes updated
4
updated fully
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!