Introduction: Various minimally invasive spine surgery (MISS) techniques have been developed with the goal of reducing approach-related soft-tissue trauma and its associated complications. However, there is still a debate on some of the potential drawbacks of MISS techniques, such as their longer operating times and increased intraoperative radiation. A solution to these disadvantages could be the implementation of new technologies, such as computer-assisted navigation (CAN) and surgical robotics. We compare the standard fluoroscopy MISS technique with our experience with time per screw and X-ray exposure for pedicle screw placement using the Brainlab Cirq passive robotic arm assistance coupled with the Brainlab Curve navigation system.

Methods: In the Cirq robot-assisted group (Group I), 109 screws were placed in 24 prospectively analyzed patients. In the fluoroscopy-guided group, 108 screws inserted into 20 consecutive patients were analyzed retrospectively (Group II). The duration of surgery, the time to place one screw, the X-ray exposition, and the pedicle screw accuracy for each patient were recorded and reviewed.

Results: In total, 217 screws were analyzed. The treated levels ranged from T10 to S1. In Group I, 104 screws were grade A (95.4%) and five were grade B (4.6%). In Group II, 96 screws were grade A (88.89%); ten were grade B (9.26%); one was grade C (0.93%), and one was grade D (0.93%). While the screws placed by using the Cirq system were more accurate overall, there was no statistical significance when the two groups were compared, p = 0.3724. There was no significant difference in radiation exposure between the two groups, p = 0.5482; however the radiation exposure for the surgeon was very limited with the Cirq system. There was a significant reduction in the operation length (p = 0.0183) and the time per screw (p < 0.0001) for Group I.

Conclusions: The CAN systems and emerging robotic platforms have the potential to diminish the main disadvantages of MISS techniques-longer operation times and X-ray exposure, at least for the surgical team.

Download full-text PDF

Source
http://dx.doi.org/10.1007/978-3-031-36084-8_59DOI Listing

Publication Analysis

Top Keywords

pedicle screw
12
screw placement
8
minimally invasive
8
invasive spine
8
spine surgery
8
time screw
8
screw x-ray
8
screws grade
8
grade 093%
8
cirq system
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!