Background: The BEBIG Portio multi-channel applicator provides better target dose coverage and sparing organs-at-risk compared to a single-channel cylinder. However, artifacts and distortions of Portio in magnetic resonance images (MRI) have not yet been reported.

Objective: We aimed to quantify the artifacts and distortions in its 1.5-Tesla MR images before clinical use.

Material And Methods: In this experimental study, we employed a gelatin-filled phantom to conduct our measurements. T2-weighted (T2W) images were examined for artifacts and distortions. Computed tomography (CT) images were used as a reference to assess image distortions. Artifact severity was measured by recording the full-width-at-half-maximum (FWHM) image pixel values at various positions along the length of the applicator/channels. CT and MRI-based applicator reconstruction accuracy were then compared, and signal-to-noise ratio (SNR) and contrast were also determined for the applicator images.

Results: The applicator distortion level for the Portio applicator was less than the image spatial resolution (0.5±0.5 pixels). The average FWHM for the tandem applicator images was 5.23±0.39 mm, while it was 3.21±0.37 mm for all channels (compared to their actual diameters of 5.0 mm and 3.0 mm, respectively). The average applicator reconstruction difference between CT and MR images was 0.75±0.30 mm overall source dwell positions. The image SNR and contrast were both acceptable.

Conclusion: These findings indicate that the Portio applicator has a satisfactory low level of artifacts and image distortions in 1.5-Tesla, T2W images. It may, therefore, be a promising option for MRI-guided multi-channel vaginal brachytherapy.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10749412PMC
http://dx.doi.org/10.31661/jbpe.v0i0.2309-1665DOI Listing

Publication Analysis

Top Keywords

image distortions
12
artifacts distortions
12
applicator
9
artifacts image
8
magnetic resonance
8
images
8
resonance images
8
multi-channel vaginal
8
distortions 15-tesla
8
t2w images
8

Similar Publications

Background: Diffusion-weighted (DW) turbo-spin-echo (TSE) imaging offers improved geometric fidelity compared to single-shot echo-planar-imaging (EPI). However, it suffers from low signal-to-noise ratio (SNR) and prolonged acquisition times, thereby restricting its applications in diagnosis and MRI-guided radiotherapy (MRgRT).

Purpose: To develop a joint k-b space reconstruction algorithm for concurrent reconstruction of DW-TSE images and the apparent diffusion coefficient (ADC) map with enhanced image quality and more accurate quantitative measurements.

View Article and Find Full Text PDF

Introduction: Vocal distortion, also known as a scream or growl, is used worldwide as an essential technique in singing, especially in rock and metal, and as an ethnic voice in Mongolian singing. However, the production mechanism of vocal distortion is not yet clearly understood owing to limited research on the behavior of the larynx, which is the source of the distorted voice.

Objectives: This study used high-speed digital imaging (HSDI) to observe the larynx of professional singers with exceptional singing skills and determine the laryngeal dynamics in the voice production of various vocal distortions.

View Article and Find Full Text PDF

False-Negative Review from the Mammography Audit: Refining Breast Imaging Practice.

Radiographics

February 2025

From the Washington University School of Medicine, Mallinckrodt Institute of Radiology, 510 S Kingshighway Blvd, St. Louis, MO 63110.

Annual review of false-negative (FN) mammograms is a mandatory and critical component of the Mammography Quality Standards Act (MQSA) annual mammography audit. FN review can help hone reading skills and improve the ability to detect cancers at mammography. Subtle architectural distortion, asymmetries (seen only on one view), small lesions, lesions with probably benign appearance (circumscribed regular borders), isolated microcalcifications, and skin thickening are the most common mammographic findings when the malignancy is visible at retrospective review of FN mammograms.

View Article and Find Full Text PDF

High-resolution awake mouse fMRI at 14 tesla.

Elife

January 2025

Athinoula A. Martinos Center for Biomedical Imaging, Department of Radiology, Harvard Medical School, Massachusetts General Hospital, Charlestown, United States.

High-resolution awake mouse functional magnetic resonance imaging (fMRI) remains challenging despite extensive efforts to address motion-induced artifacts and stress. This study introduces an implantable radio frequency (RF) surface coil design that minimizes image distortion caused by the air/tissue interface of mouse brains while simultaneously serving as a headpost for fixation during scanning. Furthermore, this study provides a thorough acclimation method used to accustom animals to the MRI environment minimizing motion-induced artifacts.

View Article and Find Full Text PDF

Background: The common marmoset (Callithrix jacchus) is an important animal model in neuroscience and neurological diseases, presenting primate-specific evolutionary features such as an expanded frontal cortex. We established a new consortium with funding support from the National Institute on Aging to generate, characterize, and validate MArmosets as Research MOdels of AD (MARMO-AD). This consortium develops and studies gene-edited marmoset models carrying genetic risk for AD, comparing them against wild-type aging marmosets from birth throughout their lifespan, using non-invasive longitudinal assessments.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!