Objectives: This in vitro study is aimed at assessing whether implant primary stability is influenced by implant length in artificial bone with varying densities.

Materials And Methods: A total of 120 truncated-conical implants (60 long-length: 3p L, 3.8 × 14 mm; 60 short-length: 3p S, 3.8 × 8 mm) were inserted into 20, 30, and 40 pounds per cubic foot (PCF) density polyurethane blocks. The insertion torque (IT), removal torque (RT), and resonance frequency analysis (RFA) values were recorded for each experimental condition.

Results: In 30 and 40 PCF blocks, 3p S implants exhibited significantly higher IT values (90 and 80 Ncm, respectively) than 3p L (85 and 50 Ncm, respectively). Similarly, RT was significantly higher for 3p S implants in 30 and 40 PCF blocks (57 and 90 Ncm, respectively). However, there were no significant differences in RFA values, except for the 20 PCF block, where 3pS implants showed significantly lower values (63 ISQ) than 3p L implants (67 ISQ) in both the distal and mesial directions.

Conclusions: These results demonstrated that the implant's length mainly influences the IT and RT values in the polyurethane blocks that mimic the mandibular region of the bone, resulting in higher values for the 3p S implants, while the RFA values remained unaffected. However, in the lowest density block simulating the maxillary bone, 3p L implants exhibited significantly higher ISQ values.

Clinical Relevance: Therefore, our data offer valuable insights into the biomechanical behavior of these implants, which could be clinically beneficial for enhancing surgical planning.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s00784-023-05444-xDOI Listing

Publication Analysis

Top Keywords

polyurethane blocks
12
rfa values
12
implant length
8
primary stability
8
vitro study
8
implants
8
pcf blocks
8
implants exhibited
8
exhibited higher
8
higher values
8

Similar Publications

Influence of Machining Parameters on the Surface Roughness and Tool Wear During Slot Milling of a Polyurethane Block.

Materials (Basel)

January 2025

Institute of Machine Tools and Production Engineering, Faculty of Mechanical Engineering, Lodz University of Technology, Stefanowskiego 1/15, 90-537 Lodz, Poland.

The aim of the work was to investigate the influence of the machining parameters on the surface roughness and tool wear during slot milling of a polyurethane block (PUB). In the experiment, the influence of the cutting speed, the feed per tooth and the depth of cut on the roughness and of the milling slot surface and wear of the end mill was analyzed. A three-axis CNC milling machine Emco Concept Mill 55 was used to perform the study.

View Article and Find Full Text PDF

A Novel Polytetrahydrofuran-Based Shape Memory Polyurethane Enhanced by Polyglycolide-Block-Polytetrahydrofuran-Block-Polyglycolide Copolymer.

Polymers (Basel)

December 2024

State Key Laboratory of Organic-Inorganic Composites, Beijing Laboratory of Biomedical Materials, College of Life Science and Technology, Beijing University of Chemical Technology, Beijing 100029, China.

A series of polyurethanes (PU-GT) were prepared using polyglycolide-block-polytetrahydrofuran-block-polyglycolide (PGA-PTHF-PGA), polytetrahydrofuran homopolymer (PTHF), glycerol, and hexamethylene diisocyanate (HDI) by a one-pot synthesis method. The non-isothermal crystallization and subsequent heating curves showed that the PTHF component in these polyurethanes could crystallize in a temperature range of -11.5~2.

View Article and Find Full Text PDF

: This study aimed to evaluate the primary stability, according to the insertion torque value (ITV) and resonance frequency analysis (RFA), of dental implants placed in standardized blocks of bone quality equivalent to type II-A bone, using three surgical undersized protocols of 0.2 mm, 0.5 mm, and 0.

View Article and Find Full Text PDF
Article Synopsis
  • This paper discusses the creation and organization of polyurethane elastomers using special bis-aromatic urea hydrogen bonding motifs to enhance self-assembly.
  • It details the use of polytetramethylene ether glycol (PTMG) and other diols to form supramolecular polyurethanes with specific molecular weights and distribution.
  • The resulting materials display adjustable mechanical properties and healing capabilities, making them suitable for practical applications like cable coatings.
View Article and Find Full Text PDF

Polyurethane (PUR) soft foams release malodorous and potentially toxic compounds when exposed to oxidative conditions. Current chamber test methods cannot distinguish between pre-existing volatiles and those formed during oxidation, nor can they assess the formation rates of oxidation products. We subjected PUR soft foam to oxidative treatment in a continuous air flow at 120 °C.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!