Gridless adult cervical spine radiography and its' effect on image quality and radiation dose: A phantom study.

Radiography (Lond)

Department of Technical Medical Radiology, School of Health Sciences (HESAV), University of Applied Sciences and Arts of Western Switzerland (HES-SO), Av. de Beaumont 21, 1011 Lausanne, Switzerland.

Published: January 2024

Introduction: Imaging of the cervical spine in general radiography is most frequently performed using an anti-scatter grid. The purpose of this study was to investigate the effects of a gridless setting on image quality and radiation dose during digital radiography of the anteroposterior (AP) and lateral (LAT) cervical spine.

Methods: A phantom study was performed with a variety of tube voltages (63-75 kV) with and without an anti-scatter grid. The tube current time product (mAs) and dose area product (DAP) were recorded and used to calculate effective dose (ED) and individual organ dose using PCXMC 2.0 software, as well as entrance surface dose (ESD) and objective image quality: signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR). Subjective visual image quality grading characteristics (VGC) was performed by five qualified radiographers.

Results: In a gridless setting, the AP and LAT positions showed significantly lower DAP (1.6 μGym; 61.3 % and 1.6 μGym; 51.2%), ESD (27.6 μGy; 57.3% and 77.2 μGy; 47.2%) and ED (4.2 μSv; 61.3% and 2.3 μSv; 48.9%). In a gridless setting in the AP position, there is a slight significant deterioration in image quality. In the lateral projection, on the other hand, the image quality without the use of grid was only significantly reduced in three of six criteria and there was no difference in the objective image quality between the two settings examined.

Conclusion: The results of this study show that gridless setting significantly decreases radiation dose and image quality, but the quality in the lateral projection is still acceptable for diagnostic purpose.

Implications For Practice: The protocol without the use of the anti-scatter grid in cervical spine radiography leads to a reduction in the radiation dose in both projections, but the image quality in the AP is significantly reduced for all criteria examined, with a slight deterioration in image quality in the lateral projection.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.radi.2023.12.009DOI Listing

Publication Analysis

Top Keywords

image quality
40
radiation dose
16
gridless setting
16
cervical spine
12
anti-scatter grid
12
quality lateral
12
lateral projection
12
quality
11
image
10
spine radiography
8

Similar Publications

Background: Knee injuries resulting in purely cartilaginous defects are rare, and controversy remains regarding the reliability of chondral-only fixation.

Purpose: To systematically review the literature for fixation methods and outcomes after primary fixation of chondral-only defects within the knee.

Study Design: Systematic review; Level of evidence, 5.

View Article and Find Full Text PDF

Optimized methods for scRNA-seq and snRNA-seq of skeletal muscle stored in nucleic acid stabilizing preservative.

Commun Biol

January 2025

Center for Research on Genomics and Global Health, National Human Genome Research Institute, National Institutes of Health, Bethesda, MD, 20892, USA.

Single cell studies have transformed our understanding of cellular heterogeneity in disease but the need for fresh starting material can be an obstacle, especially in the context of international multicenter studies and archived tissue. We developed a protocol to obtain high-quality cells and nuclei from dissected human skeletal muscle archived in the preservative Allprotect® Tissue Reagent. After fluorescent imaging microscopy confirmed intact nuclei, we performed four protocol variations that compared sequencing metrics between cells and nuclei enriched by either filtering or flow cytometry sorting.

View Article and Find Full Text PDF

Objectives: The objective of this web-based study is to analyze the attributes of bariatric surgery cases ensuing health implications. Additionally, the study seeks to delve into the factors influencing post-bariatric psychological evaluations and the impact of various bariatric surgeries on weight loss and psycho-social assessment scores for patients who had undergone bariatric surgeries within a specific bariatric surgery center in Egypt between January 2017 and January 2024.

Methods: An analytical cross-sectional study recruited 411 adults who had undergone different bariatric procedures by the same surgical team.

View Article and Find Full Text PDF

Background: Bilateral risk-reducing mastectomies (RRMs) have been proven to decrease the risk of breast cancer in patients at high risk owing to family history or having pathogenic genetic mutations. However, few resources with consolidated data have detailed the patient experience following surgery. This systematic review features patient-reported outcomes for patients with no breast cancer history in the year after their bilateral RRM.

View Article and Find Full Text PDF

Machine learning techniques for non-destructive estimation of plum fruit weight.

Sci Rep

January 2025

Crop and Horticultural Science Research Department, Mazandaran Agricultural Resources Research and Education Center, Agricultural Research, Education and Extension Organization (AREEO), Tajrish, Iran.

Plum fruit fresh weight (FW) estimation is crucial for various agricultural practices, including yield prediction, quality control, and market pricing. Traditional methods for estimating fruit weight are often destructive, time-consuming, and labor-intensive. In this study, we addressed the problem of predicting plum FW using artificial intelligence (AI) methods based on fruit dimensions.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!