Falls are a prevalent cause of injury among older people. While some wearable inertial measurement unit (IMU) sensor-based systems have been widely investigated for fall risk assessment, their reliability, validity, and identification ability in community-dwelling older people remain unclear. Therefore, this study evaluated the performance of a commercially available IMU sensor-based fall risk assessment system among 20 community-dwelling older recurrent fallers (with a history of ≥2 falls in the past 12 months) and 20 community-dwelling older non-fallers (no history of falls in the past 12 months), together with applying the clinical scale of the Mini-Balance Evaluation Systems Test (Mini-BESTest). The results show that the IMU sensor-based system exhibited a significant moderate to excellent test-retest reliability (ICC = 0.838, < 0.001), an acceptable level of internal consistency reliability (Spearman's rho = 0.471, = 0.002), an acceptable convergent validity (Cronbach's α = 0.712), and an area under the curve (AUC) value of 0.590 for the IMU sensor-based receiver-operating characteristic (ROC) curve. The findings suggest that while the evaluated IMU sensor-based system exhibited good reliability and acceptable validity, it might not be able to fully identify the recurrent fallers and non-fallers in a community-dwelling older population. Further system optimization is still needed.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10742152PMC
http://dx.doi.org/10.3390/bios13120998DOI Listing

Publication Analysis

Top Keywords

imu sensor-based
24
community-dwelling older
16
sensor-based system
12
fall risk
12
risk assessment
12
older people
12
reliability validity
8
validity identification
8
identification ability
8
inertial measurement
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!