Severity: Warning
Message: file_get_contents(https://...@pubfacts.com&api_key=b8daa3ad693db53b1410957c26c9a51b4908&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests
Filename: helpers/my_audit_helper.php
Line Number: 176
Backtrace:
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3122
Function: getPubMedXML
File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global
File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword
File: /var/www/html/index.php
Line: 316
Function: require_once
Background: Femoral stem design affects periprosthetic bone mineral density (BMD), which may impact long-term survival of cementless implants in total hip arthroplasty (THA). The aim of this study was to examine proximal femoral BMD in 3 morphologically different uncemented femoral stem designs to investigate whether any particular design resulted in better preservation of BMD.
Methods: A total of 119 patients were randomized to receive a proximally coated collarless dual-taper wedge stem, a proximally coated collarless anatomic stem, or a fully coated collarless triple-taper stem. All surgeries were performed via the posterior approach, with mobilization on the day of surgery. Dual x-ray absorptiometry scans (Lunar iDXA, GE Healthcare) assessed BMD across the 7 Gruen zones preoperatively and at 6 weeks and 2 years postoperatively; if available, the native contralateral femur was also assessed as a control. Patient-reported outcomes of pain, function, and health were also assessed at these follow-ups.
Results: Averaged across all stems, BMD increased in zones 1 (2.5%), 2 (17.1%), 3 (13.0%), 5 (10%), and 6 (17.9%) at 2 years. Greater preservation of BMD was measured on the lateral cortex (zone 2) for both the dual-taper wedge and anatomic stems (p = 0.019). The dual-taper wedge stem also demonstrated preservation of BMD in the medial calcar (zone 7), while the anatomic and triple-taper stems declined in this region; however, the difference did not reach significance (p = 0.059). Averaged across all stems, BMD decreased in the mid-diaphysis region, distal to the stem tip (zone 4). All stems performed similarly at the time of final follow-up with respect to the patient-reported outcomes.
Conclusions: This study demonstrated maintenance of femoral BMD after use of 3 different cementless femoral stem designs, with all achieving excellent improvements in patient-reported outcomes. The stems designed to load the proximal metaphyseal region resulted in higher BMD in that region. No significant stress-shielding was observed; however, longer follow-up is required to elucidate the impact of this finding on implant survivorship.
Level Of Evidence: Therapeutic Level I . See Instructions for Authors for a complete description of levels of evidence.