Objectives: The purpose of this proof-of-concept study was to investigate whether a commercially available urine dipstick may provide potentially useful information for screening plasma glucose and bilirubin in human plasma samples.

Methods: Glucose and bilirubin were assayed in 60 anonymized lithium-heparin residual plasma samples using the Roche COBAS 8000 or after pipetting 10 µL of plasma onto the pads of a commercial urine dipstick. Semiquantitative urine test results obtained with the dipstick were directly compared to paired test results obtained with COBAS.

Results: Median plasma glucose values between COBAS and dipstick were slightly different (5.8 vs. 5.6 mmol/L; p=0.040), while no significant difference was found in bilirubin values between COBAS and dipstick (11.2 vs. 8.6 μmol/L; p=0.090). The Spearman's correlation between COBAS and dipstick was 0.83 (95% CI, 0.73-0.90; p<0.001) for plasma glucose and 0.78 (95% CI, 0.66-0.87; p<0.001) for plasma bilirubin, respectively. Cumulative agreement between COBAS and dipstick was high for both glucose (88%; kappa statistic statistics, 0.75; 95% CI, 0.58-0.92; p<0.001) and bilirubin (88%; kappa statistics, 0.76; 95% CI, 0.60-0.92; p<0.001).

Conclusions: The results of this proof-of-concept study indicate that the commercial urine test strip used in our study provides acceptable performance for screening plasma glucose and bilirubin levels compared with reference laboratory assays.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10724854PMC
http://dx.doi.org/10.1515/almed-2023-0114DOI Listing

Publication Analysis

Top Keywords

urine dipstick
12
plasma glucose
12
glucose bilirubin
12
cobas dipstick
12
screening plasma
8
proof-of-concept study
8
values cobas
8
plasma
6
dipstick
6
urine
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!