Background: Exact preoperative localization is desirable to perform minimally invasive parathyroidectomy for hyperparathyroidism (HPT). This study aimed to evaluate the diagnostic values of Tc-methoxyisobutylisonitrile (Tc-MIBI) single photon emission computed tomography/computed tomography (SPECT/CT) of parathyroid glands by analyzing the relationship between lesion weight and false-negative (FN) results, as well as to explain the possible reason.

Methods: The data from 314 patients with suspected HPT who underwent Tc-MIBI SPECT/CT parathyroid imaging between 2011 and 2022 were retrospectively evaluated. The sensitivity, positive predictive value (PPV), negative predictive value (NPV), and accuracy of parathyroid Tc-MIBI SPECT/CT were calculated, and the false-positive (FP) and FN findings were analyzed.

Results: Accurate localization by Tc-MIBI SPECT/CT was significantly associated with the parathyroid hormone (PTH) level. The Tc-MIBI SPECT/CT for diagnosis/lesion location reached a sensitivity of 84.6%/56.8%, a PPV of 97.3%/98.4%, an NPV of only 23.7%/4.18%, and an accuracy of 83.4%/57.1%, respectively. The largest diameter, shortest diameter, and lesion volume were lower in the FN group than in the TP group. A total of 7 FP cases were found, including 2 cases of thyroid nodules, 4 cases of thyroid tissue, and 1 case of hibernoma. A total of 45 FN patients, including 321 FN lesions, were confirmed, of which parathyroid hyperplasia accounted for 97.8%. Lesion weights greater than 20 µg were able to be detected, but lightweight lesions less than 100 mg were the principal source of FN results, accounting for approximately 39.3%. With lesion weights 0-100, 101-300, 301-1,000, and >1,000 mg, the FN rate was 70.8% (126/178), 51.8% (103/199), 34.6% (81/234), and 8.33% (11/132), respectively.

Conclusions: Tc-MIBI SPECT/CT parathyroid imaging provides good sensitivity and high specificity in HPT location. Correct localization by Tc-MIBI SPECT/CT correlates positively with lesion weight and PTH levels. The smaller the lesion, the higher the FN rate in Tc-MIBI SPECT/CT parathyroid imaging, and lesions weighing less than 100 mg are the main source of FN results in Tc-MIBI SPECT/CT parathyroid imaging.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10722006PMC
http://dx.doi.org/10.21037/qims-23-1091DOI Listing

Publication Analysis

Top Keywords

tc-mibi spect/ct
36
spect/ct parathyroid
24
parathyroid imaging
20
tc-mibi
10
spect/ct
10
parathyroid
9
lesion weight
8
localization tc-mibi
8
cases thyroid
8
lesion weights
8

Similar Publications

Localization in primary hyperparathyroidism.

Best Pract Res Clin Endocrinol Metab

December 2024

Department of Nuclear Medicine, Post Graduate Institute of Medical Education and Research, Sector 12, Chandigarh 160012, India. Electronic address:

Primary hyperparathyroidism is the main cause of hypercalcemia, resulting predominantly from parathyroid adenomas followed by hyperplasia. Diagnosis relies on clinical and biochemical parameters. Accurate pre-operative localization is mandatory for better surgical outcome.

View Article and Find Full Text PDF
Article Synopsis
  • - Minimally invasive parathyroidectomy is the preferred treatment for primary hyperparathyroidism, but it relies on effective preoperative localization methods to identify overactive parathyroid tissue.
  • - A study compared two groups of patients—234 who had traditional MIBI scintigraphy and 163 who had FCh-PET/CT—focusing on the efficiency and outcomes of these imaging techniques leading up to surgery.
  • - Results showed that FCh-PET/CT significantly reduced imaging time by 83% and surgery time by 41% without using intraoperative hormone monitoring, suggesting it may lead to faster procedures and fewer reoperations for patients.
View Article and Find Full Text PDF
Article Synopsis
  • This study assesses the effectiveness of prone SPECT imaging with [Tc]Tc-sestamibi compared to Molecular Breast Imaging (MBI) in predicting treatment response in women with locally advanced breast cancer (LABC).
  • It involved 18 patients who underwent SPECT scans shortly after and 90 minutes post-injection, along with MBI, to analyze tumor characteristics like wash-out rates and standardized uptake values.
  • Results indicated that early SPECT provided useful semi-quantitative data comparable to MBI, though further research with broader clinical parameters is necessary to determine the clinical implications of these findings.
View Article and Find Full Text PDF

Purpose: Parathyroid imaging with dual-phase technetium-99m methoxyisobutrylizonitrile (Tc-99m MIBI) scintigraphy serves as an important prerequisite for the identification of hyperfunctioning parathyroid gland(s) in patients with primary hyperparathyroidism (PHPT) for a successful targeted parathyroidectomy. This study aimed to evaluate the clinical value of additional lateral imaging and single-photon emission computed tomography/computed tomography (SPECT/CT) versus conventional planar imaging for locating parathyroid pathologies in patients with PHPT.

Materials And Methods: A retrospective review was performed on 105 patients who underwent dual-phase Tc-99m MIBI scintigraphy and were surgically treated by parathyroidectomy.

View Article and Find Full Text PDF

Impact of CT attenuation correction on viable myocardium detection in combined SPECT and PET/CT: A retrospective cohort study.

Medicine (Baltimore)

October 2024

Department of Nuclear Medicine, TEDA International Cardiovascular Hospital, Tianjin University, Tianjin, P.R. China.

The influence of computed tomography attenuation correction (CTAC) on the accuracy of diagnosing viable myocardium using Tc-99m-MIBI dedicated cardiac cadmium-zinc-telluride (CZT) single-photon emission computed tomography (SPECT) myocardial perfusion imaging (MPI) combined with F-18-FDG Positron Emission Tomography/Computed Tomography (PET/CT) metabolic imaging, compared with conventional SPECT MPI, remains to be fully elucidated. To evaluate the impact of CTAC on the accuracy of diagnosing viable myocardium using Tc-99m-MIBI dedicated cardiac CZT SPECT MPI combined with F-18-FDG PET/CT, compared to conventional SPECT MPI. 193 patients underwent CZT SPECT and F-18-FDG PET/CT imaging, while 39 patients underwent conventional SPECT and F-18-FDG PET/CT imaging, with both groups utilizing CT for attenuation correction.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!