Article Abstract

Introduction: The aim of the current study was to elucidated the reliable atrial fibrillation (AF) drivers identified by CARTOFINDER using OctaRay catheter.

Methods And Results: The reliability of focal and rotational activations identified by CARTOFINDER using OctaRay catheter was assessed by the sequential recordings in each site of both atrium before and after pulmonary vein isolation (PVI) in 10 persistent AF patients. The outcome measures were the reproducibility rate during the sequential recordings and the stability rate between pre- and post-PVI as reliable focal and rotational activations. The study results were compared with those under use of PentaRay catheter (N = 18). Total 68928 points of 360 sites in OctaRay group and 24 177 points of 311 sites in PentaRay were assessed. More focal activation sites were identified in OctaRay group than PentaRay group (7.9% vs. 5.7%, p < .001), although the reproducibility rate and the stability rate were significantly lower in OctaRay group (45.3% vs. 58.9%, p < .001; 11.2% vs. 28.4%, p < .001). Meanwhile, the prevalence of reproducible focal activation sites among overall points was comparable (3.6% vs. 3.3%, p = .08). Regarding rotational activation, more rotational activation sites were identified in OctaRay group (5.1% vs. 0.2%, p < .001), and the reproducibility rate and the stability rate were significantly higher in OctaRay group (45.2% and 12.5% vs. 0.0%, p < .001). Both reliable focal and rotational activation sites were characterized by significantly shorter AF-cycle length (CL) and higher repetition of focal and rotational activations during the recordings compared with the sites of non or unreliable focal and rotational activations.

Conclusion: In CARTOFINDER, OctaRay catheter could identify reliable focal activation with high resolution and reliable rotational activation compared with PentaRay catheter. The repetitive focal and rotational activations with short AF-CL could be the potential target during ablation.

Download full-text PDF

Source
http://dx.doi.org/10.1111/jce.16159DOI Listing

Publication Analysis

Top Keywords

focal rotational
12
rotational activations
12
reliable focal
8
octaray catheter
8
identified cartofinder
8
cartofinder octaray
8
sequential recordings
8
octaray group
8
octaray
5
activations cartofinder
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!