Purpose: RigidCare is an electrolysis-based device that recently obtained approval from the US's FDA to sterilise microorganisms and remove proteins for orthokeratology (O-K) lenses. The study was conducted to investigate the device's performance in varied clinical circumstances.
Methods: Trial lenses and private lenses were employed by O-K lens wearers from five hospitals for an evaluation of disinfection and sterilisation and an assessment of protein removal, respectively. Menicon multipurpose solution and protein remover were selected for use with the control group. Following the instructions, pre-cleaning lens samples, post-cleaning lens samples and residual solution samples of trial lenses of the experimental and control groups were collected for microorganism examinations by an experienced third-party testing organisation. The levels of protein deposition for these two approaches were rated by senior O-K experts. Categorical variables were analysed using statistical tests, such as the chi-squared test and Fisher's exact test.
Results: The microbial positive rate detected from the pre-cleaning and post-cleaning lens samples and the residual solution of the trial lenses for the experimental and control group was 4/76 vs 1/74 (P = 0.37), 1/76 vs 0/74 (P = 1.00) and 0/76 vs 8/74 (P = 0.006), respectively. Following protein removal, the experimental group exhibited a significantly higher overall proportion of lenses rated as 'clean' or with a 'mild deposit' (96.4 %, 79/82) compared to the control group (85.7 %, 66/77), with a significant difference (P < 0.05).
Conclusion: This multi-center study demonstrated that RigidCare exhibited superior efficacy in disinfection, sterilisation and protein removal as compared to Menicon multipurpose solution and protein remover.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.clae.2023.102106 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!