Study of the position of six volar locking plates for distal radius fracture relative to the watershed line.

Hand Surg Rehabil

Service de Chirurgie Orthopédique et Traumatologique, Hôpital de la Cavale Blanche, Boulevard Tanguy Prigent, 29200 Brest, France; Université de Bretagne Occidentale, UBO, 29200 Brest, France; LaTIM, INSERM, UMR 1101, SFR IBSAM, Avenue Foch, 29200 Brest, France. Electronic address:

Published: February 2024

Introduction: Volar plate malpositioning in the treatment of distal radial fracture can lead to tendinitis or even tendon tear, especially when the plate position is very distal. We studied the impact of design on plate position in the distal radius. The primary aim was to compare the position of six volar wrist plates relative to the watershed line using the Soong classification. The secondary objectives were to assess the epidemiology of volar locking plate fixation within the administrative Département of Finistère (northwestern France) and to study whether the type of fracture played a role in plate position.

Hypothesis: The plate design itself influences positioning relative to the watershed line on the Soong classification.

Materials And Methods: A total of 2723 volar locking plate fixation cases were analyzed and categorized according to the Soong classification. Plates used were divided into six groups based on design: Zimmer Biomet®, Newclip Technics®, Stryker®, Synthes®, Medartis® and Medartis® Footprint. The number of Soong 0 + 1 plates (i.e., plates graded 0 and 1 taken together) was determined for each design, then compared using the Marascuilo procedure with a significance level of α = 0.05.

Results: On the Marascuilo procedure, we found significant differences in the number of Soong grade 0 + 1 plates. The Zimmer Biomet and Newclip Technics® plates were significantly more often proximal to the watershed line than the Synthes and Medartis Footprint plates. Plate position with the Medartis® design was significantly more proximal to the watershed line than for its companion design, the Medartis® Footprint plate. The rate of volar locking plate fixation of distal radial fractures over the past 10 years increased in Finistère. Also, the type of fracture affected the choice of plate when different designs were available within a hospital center (Medartis® Footprint plate used in 2R3A fractures).

Conclusion: Our study highlights a significant difference in volar locking plate position relative to the watershed line between the various models available. Plate design is a deciding factor when treating distal radial fracture, to avoid impingement when implant removal is not routinely planned.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.hansur.2023.11.009DOI Listing

Publication Analysis

Top Keywords

volar locking
20
relative watershed
16
plate position
16
locking plate
16
plate
14
distal radial
12
plate fixation
12
medartis® footprint
12
position volar
8
plates
8

Similar Publications

There are many approaches to the wrist both volar and dorsal, depending on the injury at hand. The design of the volar locking plate has created a rise in distal radius fractures being treated using a volar FCR approach and its modifications. It does, however, have limitations in visualization of the volar ulnar corner of the radius.

View Article and Find Full Text PDF

Treatment of anterior rim fractures of the distal radius with an anterior rim plate: A retrospective study.

J Hand Surg Eur Vol

January 2025

Department of Orthopedics, E-Da Hospital, I-Shou University/School of Medicine, College of Medicine, I-Shou University, Kaohsiung, Taiwan.

Treatment of distal radial fractures that include small anterior rim fragments can be difficult. We retrospectively reviewed 19 patients in whom an anterior rim plate with locking screws was used. After a median follow-up of 18 months (range 6-32; interquartile range (IQR) 14, 26), the median wrist flexion and extension arc was 70° (range 50-80; IQR 60, 70), the median grip strength was 80% of the contralateral side (range 52-104; IQR 77, 88), the median visual analogue scale score for pain was 0 (range 0-5; IQR 0, 1), the median disabilities of the arm, shoulder and hand score was 2 (range 0-59; IQR 0, 11) and the median modified Mayo wrist score was 80 (range 35-100; IQR 75, 85).

View Article and Find Full Text PDF

Background: Extra-articular but severely comminuted distal basal fractures of the proximal phalanx (PP) are rarely reported. Therefore, the aim of this study was to achieve proper union and desirable outcomes using low-profile locking plates/screws. We introduced our own surgical approach and reported the clinical/radiographic outcomes via retrospective case series.

View Article and Find Full Text PDF

Risk factors for complications following volar locking plate fixation of distal radial fractures.

Orthop Traumatol Surg Res

December 2024

Service de Chirurgie Orthopédique et Traumatologique, Hôpital de la Cavale Blanche, Boulevard Tanguy Prigent, Brest 29200, France; Université de Bretagne Occidentale, UBO, Brest 29200, France; LaTIM, INSERM, UMR 1101, SFR IBSAM, Avenue Foch, Brest 29200, France. Electronic address:

Article Synopsis
  • The study focuses on distal radial fractures and the complications associated with volar locking plate (VLP) fixation, which can lead to implant removal.
  • Researchers analyzed data from 2951 patients, identifying intrinsic (like age and fracture type) and extrinsic (such as plate positioning) risk factors for complications post-fixation.
  • Key findings revealed that being under 62 years old, having a specific type of fracture, and the plate's positioning were significant risk factors for complications that might necessitate implant removal.
View Article and Find Full Text PDF
Article Synopsis
  • The study investigates the effectiveness of two types of bone grafts (packed cancellous vs. corticocancellous) used in repairing scaphoid waist nonunion fractures with volar locking plates.
  • Postoperative CT scans showed that 95% of patients achieved osseous bridging, but those with packed cancellous grafts had significantly better healing outcomes than those with corticocancellous grafts.
  • The findings suggest that using packed cancellous autografts may lead to faster and more effective bone healing in these cases.
View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!