Spectroscopy is utilised extensively in medical sensing technology. Typically, hand-held spectroscopy equipment uses miniature narrow-band light emitting diodes (LEDs) and photodiodes to emit and detect light, respectively. Photodiodes typically absorb light across a wide spectra so measurements can be corrupted by surrounding light. LEDs in the visible spectrum have a narrower spectral response and can be used in place of a traditional photodiode. However, the absorption characteristics of near infrared (NIR) spectrum LEDs is unknown. A discrete, low-cost spectrophotometer was designed to assess spectral response for 8 narrow band NIR LEDs. The normalised and raw spectral response determined the optimum detector for 1050 nm - 1300 nm is the 1450 nm LED, and the optimum detector for 1450 nm - 1650 nm emissions is the 1650 nm LED.Clinical relevance - Understanding the spectral response of narrow-band LEDs in the NIR spectrum will aid development of NIR hand-held spectroscopy medical devices.