Purpose: This study aimed to determine the characteristics of 2D ionization chamber array and the confidence limits of the gamma passing rate in pencil beam scanning proton therapy.

Materials And Methods: The Varian ProBeam Compact spot-scanning system and the PTW OCTAVIUS 1500XDR array were used as a proton therapy system and detector, respectively. Our methods consisted of 2 parts: (1) the characteristics of the detector were tested and (2) patient-specific quality assurance was performed and evaluated by gamma analysis using dose-difference and distance-to-agreement criteria of 3% and 2 mm, respectively, with 123 treatment plans in head and neck, breast, chest, abdomen, and pelvic regions.

Results: The PTW OCTAVIUS 1500XDR array had good reproducibility, uniformity, linearity, repetition rate, and monitor unit per spot within 0.1%, with accuracy, energy dependence, and measurement depth within 0.5%. The overall uncertainty of the PTW OCTAVIUS 1500XDR array was 2.49%. For field size and range shifter, using gamma analysis, the passing rate was 100%. The overall results of patient-specific quality assurance with the gamma evaluation were 98.9% ± 1.6% in 123 plans and confidence limit was 95.7%.

Conclusion: The PTW OTAVIUS 1500XDR offered effective performance in pencil beam scanning proton therapy.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10698631PMC
http://dx.doi.org/10.14338/IJPT-23-00016.1DOI Listing

Publication Analysis

Top Keywords

patient-specific quality
12
quality assurance
12
pencil beam
12
beam scanning
12
ptw octavius
12
octavius 1500xdr
12
1500xdr array
12
passing rate
8
scanning proton
8
proton therapy
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!