In the present work, potentiometric sensors with polymer membranes used for chlorhexidine (CHXD) determination were developed. The polymer membranes were plasticized with bis(2-ethylheksyl)sebacate (DOS) or 2-nitrophenyloctyl ether (-NPOE). The active compounds used in the membrane were cyclodextrins, crown ethers, and ion exchangers. The best-constructed electrode was based on neutral heptakis(2,3,6-tri-O-benzoyl)--cyclodextrin with lipophilic salt (KTpClBP)-potassium tetrakis(4-chlorophenyl) borate-dissolved in plasticizer, DOS. The presented electrode is characterized by an average cationic slope of 30.9 ± 2.9 mV decade within a linear range of 1 × 10 to 1 × 10 mol × L, while the value of the correlation coefficient is 0.9970 ± 0.0026. The response time was about 5 s when increasing the sample concentration and about 10 s when diluting the sample. The electrode potential is independent of the pH within a range of 4.0-9.5. The polymeric membrane sensor was successfully applied for assays of chlorhexidine digluconate in pure samples and pharmaceutical samples. The relative error from three replicate measurements was determined to be 1.1%. and the accuracy was RSD = 0.3-1.1%.