Application of O-RADS US combined with MV-Flow to diagnose ovarian-adnexal tumors.

Ultrasonography

Obstetrics and Gynecology Ultrasound Department, First Affiliated Hospital of Xinjiang Medical University, Xinjiang Key Laboratory of Ultrasound Medicine, Urumqi, China.

Published: January 2024

Purpose: This study aimed to explore the application of Ovarian-Adnexal Reporting and Data System Ultrasound (O-RADS US) combined with MV-Flow (Samsung Medison Co., Ltd.) to diagnose ovarian-adnexal masses.

Methods: A total of 112 ovarian-adnexal masses (81 benign and 31 malignant) from 105 consecutive patients were analyzed. The O-RADS US and vascular index from MV-Flow (VIMV) were measured and compared with the reference standard. O-RADS US and MV-Flow were tested for consistency.

Results: Receiver operating characteristic curves were drawn for O-RADS US, MV-Flow, and their combination. The combined methods had the largest area under the curve (0.955), followed by O-RADS US (0.929) and MV-Flow (0.923). A mass was considered malignant when the O-RADS US classification was 5 and VIMV was ≥7.15. With this definition, MV-Flow had the highest sensitivity (87.10%), with consistent findings for the combined diagnostic methods and O-RADS US (83.87%). The specificity of the combined diagnostic methods (93.83%) was higher than that of MV-Flow (91.36%). O-RADS US had the lowest specificity (90.12%). The combined diagnostic methods had the highest coincidence rate (91.07%), and MV-Flow (90.18%) had a significantly higher coincidence rate than O-RADS US (88.39%). Both O-RADS US and MV-Flow showed good consistency among different physicians (former kappa, 0.974; latter intraclass correlation coefficient [ICC], 0.986). MV-Flow had a high consistency for the same physician (ICC, 1).

Conclusion: O-RADS US and MV-Flow exhibited good diagnostic efficacy, and their combined diagnostic efficacy was higher than that of each individually. O-RADS US and MV-Flow can improve the diagnosis of benign and malignant ovarian-adnexal masses.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10766884PMC
http://dx.doi.org/10.14366/usg.23061DOI Listing

Publication Analysis

Top Keywords

o-rads mv-flow
20
combined diagnostic
16
mv-flow
13
o-rads
12
diagnostic methods
12
o-rads combined
8
combined mv-flow
8
diagnose ovarian-adnexal
8
ovarian-adnexal masses
8
benign malignant
8

Similar Publications

Application of O-RADS US combined with MV-Flow to diagnose ovarian-adnexal tumors.

Ultrasonography

January 2024

Obstetrics and Gynecology Ultrasound Department, First Affiliated Hospital of Xinjiang Medical University, Xinjiang Key Laboratory of Ultrasound Medicine, Urumqi, China.

Purpose: This study aimed to explore the application of Ovarian-Adnexal Reporting and Data System Ultrasound (O-RADS US) combined with MV-Flow (Samsung Medison Co., Ltd.) to diagnose ovarian-adnexal masses.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!