This paper reports on the development of a novel soft robotic system for remote ultrasound applications. Direct contact of the ultrasound probe with the patient's body represents a safety risk and therefore control of the probe's positioning and applied force is a crucial task. The proposed robot uses a passive control system that provides safe interaction between the robot and the patient by leveraging soft robotics technology. The soft robot's structure can be considered as a nonlinear spring which can be designed to exert a safe force within the robot's workspace to guarantee the safety of human-robot interaction. The literature suggests that effective ultrasound imaging of both the heart and abdomen requires six degrees-of-freedom. These degrees-of-freedom consist of three translational motions, which are achieved using a novel hybrid soft cable-driven parallel robot, and three wrist motions, which is based on a universal joint design. The experimental results show that the robot can achieve all these six degrees-of-freedom, and its blocking force can be engineered to generate a uniform force within the workspace.