A PHP Error was encountered

Severity: Warning

Message: file_get_contents(https://...@pubfacts.com&api_key=b8daa3ad693db53b1410957c26c9a51b4908&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests

Filename: helpers/my_audit_helper.php

Line Number: 176

Backtrace:

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3122
Function: getPubMedXML

File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global

File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword

File: /var/www/html/index.php
Line: 316
Function: require_once

Dual-energy index variation when evaluating the potential ferromagnetism of ex vivo bullets. | LitMetric

Dual-energy index variation when evaluating the potential ferromagnetism of ex vivo bullets.

SA J Radiol

Department of Clinical Imaging Sciences, Faculty of Health Sciences, University of the Free State, Bloemfontein, South Africa.

Published: July 2023

Background: An MRI is potentially hazardous for patients with retained ferromagnetic bullets. Recent studies have aimed to develop dual-energy computed tomography (DECT) as a screening tool for recognising highly ferromagnetic bullets. Inconsistent findings have been ascribed to inherent CT technology differences. Previous research demonstrated significant Hounsfield unit (HU) measurement variation among single-source CT machines.

Objectives: This study investigated the theoretical dual-energy index (DEI) variation between DECT machines when evaluating the potential ferromagnetic properties within the same sample of ex vivo bullets and metal phantoms.

Method: An experimental ex vivo study was conducted on eight metal phantoms and 10 unused bullets individually positioned in the same Perspex head phantom and scanned on two DECT machines. Two senior radiology registrars independently recorded the HU readings, and DEI values were calculated. Statistical analysis was performed using non-parametric methods for paired data, namely the Signed Rank Test. The DEI values based on mean HU readings between the DECT machines were compared.

Results: Inter- and intra-reader agreement was not statistically significant. The metal phantoms had poor interscanner agreement, with an overlap of the ferromagnetic and non-ferromagnetic ranges. The bullets had good interscanner agreement, with a similar ferromagnetic to non-ferromagnetic relationship.

Conclusion: The use of DEI values negates the previous assumption that significant interscanner variability exists among different DECT technologies while assessing highly attenuative ex vivo bullets.

Contribution: This investigation demonstrated that even though HU readings may be variable, the implementation of the DEI equation translates this into comparable values with good interscanner agreement.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10696535PMC
http://dx.doi.org/10.4102/sajr.v27i1.2701DOI Listing

Publication Analysis

Top Keywords

dect machines
12
dei values
12
interscanner agreement
12
evaluating potential
8
vivo bullets
8
ferromagnetic bullets
8
metal phantoms
8
ferromagnetic non-ferromagnetic
8
good interscanner
8
bullets
6

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!