Purpose: To collate the accuracy of two recently introduced intraocular lens (IOL) formulas (VRF and VRF-G) in cataract patients using a swept-source optical coherence tomography (SS-OCT) biometry (IOL Master 700, Carl Zeiss Meditec AG, Jena, Germany).

Patients And Methods: Data records of 295 eyes from 295 patients were included in this scrutiny. The IOLMaster 700 SS-OCT biometer was used for biometric measurements. The VRF and VRF-G formulas were compared with seven 3rd and 4th generation thin and thick-lens formulas: Haigis, Hoffer Q, Holladay 1, Holladay 2, SRK/T, T2, and Barrett Universal II. With optimized lens constants, the mean prediction error (PE) and its standard deviation (SD), the median absolute error (MedAE), the mean absolute error (MAE), and the percentage of eyes with PEs within ±0.25 D, ±0.50 D, ±0.75 D, ±1.00 D, and <±2.00 D were analyzed.

Results: Statistically significant differences were found between formulas in the whole group (Friedman test, P = 0.000). The VRF-G and Haigis formulas showed the lowest SD values (0.464 D and 0.466 D respectively). The VRF and Barrett Universal II formulas were less predictable (SD 0.471 D and SD 0.474 D respectively). The biggest proportion of eyes within ±0.50 D was found with VRF-G (76.27%), Haigis (75.59%), VRF (74.92%), and Barrett Universal II (74.92%) formulas.

Conclusion: Based on data achieved from the SS-OCT biometry, the VRF-G and Haigis methods were the more precise predictors of postoperative refraction with the biggest proportion of eyes within ±0.50 D.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10693751PMC
http://dx.doi.org/10.2147/OPTH.S439287DOI Listing

Publication Analysis

Top Keywords

vrf vrf-g
12
intraocular lens
8
swept-source optical
8
optical coherence
8
coherence tomography
8
absolute error
8
accuracy vrf
4
vrf-g intraocular
4
lens power
4
power calculation
4

Similar Publications

Purpose: To compare the accuracy of 16 intraocular lens (IOL) power calculation formulas in pediatric cataract eyes.

Patients And Methods: The data records of pediatric patients who had been implanted with three IOL models (SA60AT, MA60AC, and enVista-MX60) between 2012 and 2018 were analyzed. The accuracy of 16 IOL power calculation methods was evaluated: Barrett Universal II (BUII), Castrop, EVO 2.

View Article and Find Full Text PDF

Background: To investigate the accuracy of the VRF-L and VRF-GL IOL power calculation methods in cataract surgery after radial keratotomy (RK).

Methods: The VRF-L and VRF-GL methods were collated with nine formulas: Barrett True K (No History), Haigis, Hoffer Q, Hoffer Q (Double-K), Holladay 1 (Double-K), Holladay 2 (Double-K), PEARL-DGS (RK), SRK/T (Double-K), and T2 (Double-K). With SS-OCT biometry (IOLMaster 700, Carl Zeiss Meditec), data of 78 eyes from 78 patients with previous RK was included.

View Article and Find Full Text PDF

Introduction: The aim of this work is to compare 20 intraocular lens (IOL) power calculation formulas in medium-long eyes (24.50-25.99 mm) in terms of root mean square absolute error (RMSAE), median absolute error (MedAE), and percentage of eyes with prediction error (PE) within ± 0.

View Article and Find Full Text PDF

Accuracy of Modern Intraocular Lens Formulas in Highly Myopic Eyes Implanted With Plate-Haptic Intraocular Lenses.

Am J Ophthalmol

September 2024

From the National Clinical Research Center for Ocular Diseases (E.M., Z.C., K.F., Z.Z., J.X., C.Z., P.C., J.L., Y.E.Z.), Eye Hospital, Wenzhou Medical University, Wenzhou, China; Eye Hospital and School of Ophthalmology and Optometry (E.M., Z.C., K.F., Z.Z., J.X., C.Z., P.C., J.L., Y.E.Z.), Wenzhou Medical University, Wenzhou, Zhejiang, China; Eye Hospital of Wenzhou Medical University Hangzhou Branch (P.C., Y.E.Z.), Hangzhou, Zhejiang, China. Electronic address:

Purpose: To evaluate the predictive accuracy of modern intraocular lens (IOL) formulas and axial length (AL) adjusted traditional IOL formulas, including Wang-Koch and Cooke-modified AL (CMAL) method, in long eyes with plate-haptic IOLs, and to compare refractive prediction error variances with C-loop IOLs.

Design: Retrospective consecutive case series study.

Methods: Data from 391 eyes with Zeiss 509 M and 302 eyes with Alcon SN6CWS implants in highly myopic patients, following cataract surgery from January 2019 to November 2023, were collected.

View Article and Find Full Text PDF

The effect of corneal power on the accuracy of 14 IOL power formulas.

BMC Ophthalmol

March 2024

National Clinical Research Center for Ocular Diseases, Eye Hospital, Wenzhou Medical University, Wenzhou, 325027, China.

Background: This study evaluates the impact of corneal power on the accuracy of 14 newer intraocular lens (IOL) calculation formulas in cataract surgery. The aim is to assess how these formulas perform across different corneal curvature ranges, thereby guiding more precise IOL selection.

Methods: In this retrospective case series, 336 eyes from 336 patients who underwent cataract surgery were studied.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!