Background: VersaCross is a novel radiofrequency transseptal solution that may improve the efficiency and workflow of transseptal puncture (TSP). The aim of this study was to compare the VersaCross transseptal system with mechanical needle systems during mitral transcatheter edge-to-edge repair (M-TEER) with the PASCAL device.
Methods: This is a single-center retrospective study of consecutive patients who underwent M-TEER with the PASCAL. Transseptal puncture was undertaken with either a mechanical needle or the VersaCross wire. The primary endpoints were success of TSP and successful delivery of the Edwards sheath on the chosen delivery wire. Secondary endpoints included number of wires used, tamponade rate, interval from femoral venous access to TSP and first PASCAL device deployment, procedural death, and stroke.
Results: Thirty-three consecutive patients (10 with mechanical needle, 23 with VersaCross) who underwent M-TEER with the Edwards PASCAL device were identified. All patients had successful TSP. In the mechanical needle group, the Edwards sheath was successfully delivered on the Superstiff Amplatz wire in all cases. In the VersaCross arm, the radiofrequency wire was used successfully for delivery of the sheath in all cases. There were no cases of pericardial effusion/tamponade in either arm. Interval from femoral venous access to TSP and to deployment of the first PASCAL device was shorter with the VersaCross system. Significantly fewer wires were used with VersaCross. There were no procedural deaths or strokes in either group.
Conclusions: VersaCross appears a safe and effective method of TSP and for delivery of the 22Fr sheath for M-TEER with PASCAL.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10692349 | PMC |
http://dx.doi.org/10.1016/j.shj.2023.100203 | DOI Listing |
Sci Rep
December 2024
School of Mechanical Engineering, Qilu University of Technology (Shandong Academy of Sciences), Jinan, 250353, China.
Microtextured microneedles are tiny needle-like structures with micron-scale microtextures, and the drugs stored in the microtextures can be released after entering the skin to achieve the effect of precise drug delivery. In this study, the skin substitution model of Ogden's hyperelastic model and the microneedle array and microtexture models with different geometrical parameters were selected to simulate and analyse the flow of the microtexture microneedle arrays penetrating the skin by the finite-element method, and the length of the microneedles was determined to be 200 μm, the width 160 μm, and the value of the gaps was determined to be 420 μm. A four-pronged cone was chosen as the shape of microneedles, and a rectangle was chosen as the shape of the drug-carrying microneedle.
View Article and Find Full Text PDFZhongguo Gu Shang
December 2024
Department of Traumatic Joints, The Third Affiliated Hospital of Beijing University of Chinese Medicine, Beijing 100102, China.
Objective: To explore weight-bearing stability of Pilon fracture fixed by external fixator.
Methods: Six ankle bone models (right side) and 4 pairs (8 ankle cadaver specimens) were selected. Pilon fracture model was prepared by using the preset osteotomy line based on Ruedi Allgower Pilon fracture type.
Dent Med Probl
December 2024
Department of Prosthetic Dentistry and Gerodontology, Poznan University of Medical Sciences, Poland.
Background: Ceramic endosseous implant coatings have gained esteem due to their favorable osteoinductive and osteoconductive properties. However, such a layer may be prone to failure under in vivo conditions, which necessitates its modification.
Objectives: The aim of the present study was to modify an electrodeposited hydroxyapatite (HA) coating on titanium (Ti) with ultrashort-pulsed lasers for the incorporation of the ceramic into the sample surface and the texturing of the metal surface.
J Cancer Biol Res
February 2024
Department of Mechanical and Aerospace Engineering, University of Central Florida, USA.
The examination of the prostate biopsy procedure is essential in the optimization of the diagnostic pathway of such a prevalent affliction as prostate cancer among men worldwide. With the core needle biopsy being the standard of care for the diagnosis of prostate cancer, the ability to obtain quality core samples is directly related to patient treatment and diagnostic reliability. Needle deflection and dynamic tissue deformation are two chief sources of unrepresentative samples outside of human error.
View Article and Find Full Text PDFJ Clin Med
December 2024
Radiology and Medical Imaging University Clinic, Department XV, 'Victor Babes' University of Medicine and Pharmacy Timisoara, Eftimie Murgu Square No. 2, 300041 Timisoara, Romania.
CT-guided transthoracic core-needle biopsy (CT-TTNB) is a minimally invasive procedure that plays a crucial role in diagnosing pulmonary nodules. With high diagnostic yield and low complication rates, CT-TTNB is favored over traditional surgical biopsies, providing accuracy in detecting both malignant and benign conditions. This literature review aims to present a comprehensive overview of CT-TTNB, focusing on its indications, procedural techniques, diagnostic yield, and safety considerations.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!