Severity: Warning
Message: file_get_contents(https://...@pubfacts.com&api_key=b8daa3ad693db53b1410957c26c9a51b4908&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests
Filename: helpers/my_audit_helper.php
Line Number: 176
Backtrace:
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3122
Function: getPubMedXML
File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global
File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword
File: /var/www/html/index.php
Line: 316
Function: require_once
Objectives: Accuracy of estradiol measurements is important but conventional proficiency testing (PT) cannot assess accuracy when possibly non-commutable samples are used and method peer-group means are the targets. Accuracy-based assessment of estradiol measurements is needed.
Design And Methods: Five serum samples were prepared from single donors, frozen, and distributed overnight to 76 New York State Department of Health (NYSDOH)-certified laboratories. Participants analyzed samples for estradiol. The biases of group means were assessed against the Centers for Disease Control and Prevention (CDC)-defined targets, evaluated using the Hormone Standardization Program (HoSt) E2 performance criterion of ±12.5 %. Each laboratory's performance was evaluated using total allowable error (acceptance limits) of target ±25 % or ±15 pg/mL (55 pmol/L) (whichever was greater, NYSDOH), target ±30 % (Clinical Laboratory Improvement Amendments [CLIA]), and target ±26 % (minimal limit based on biological variation [BV]).
Results: The biases (range) were 34 % (-17 % to 175 %), 40 % (-33 % to 386 %), 16 % (-45 % to 193 %), 5 % (-27 % to 117 %), and -4% (-31 % to 21 %), for samples at estradiol of 24.1, 28.4, 61.7, 94.1, and 127 pg/mL, or 89, 104, 227, 345, and 466 pmol/L, respectively. Large positive method/analytical systematic biases were revealed for 9 commonly used method/analytical systems in the United States at low estradiol concentrations. Of the 9 analytical systems, 0, 0, 3, 7 and 6 met the HoSt criterion for the samples with estradiol at the five respective concentrations. PT evaluation showed that 59 %, 69 % and 87 % of laboratories would receive a PT event passing (satisfactory) score when the CDC-defined target and a criterion of NYSDOH, CLIA or BV was used, respectively. However, >95 % laboratories would obtain PT passing score if method peer-group means were used as targets regardless of the criterion used.
Conclusions: Improvement in accuracy of estradiol measurements is needed, particularly at low estradiol concentrations. Accuracy-based PT provides unambiguous information about the accuracy of methods/analytical systems.