Aim: High-intensity focused ultrasound (HIFU) is a novel minimally invasive local treatment of solid tumors. Endoscopic ultrasound-guided HIFU (EUS-HIFU) using mechanical effects would have potential benefits, including precise detection of target lesions and enhance drug delivery. The aim of this study is to develop EUS-HIFU device and to prove our concept in porcine model using a locally injected phase change nano droplet (PCND) as the sensitizer.

Method: A phospholipid PCND contained volatile perfluoro-carbon liquids. The prototype HIFU apparatus comprised a small (20 × 20 mm) transducer with center frequency of 2.1 MHz, attachable to a linear EUS transducer. Under general anesthetic, a single porcine received EUS-guided injection of PCND. The HIFU transducer was placed laparotomically in the stomach, and the liver was ablated through the gastric wall.

Results: PCND was injected successfully and a distinct lesion was generated at the HIFU transducer focus only in injected areas that received HIFU exposure at 4.7 kW/cm at a duty cycle of 5 % (mean temporal intensity, 0.245 kW/cm) for 30 s. The generated lesions were mechanically fractionated in macroscopic view.

Conclusion: The concept of transluminal HIFU ablation using novel EUS-HIFU system was proved in a porcine animal model. This novel treatment system has great potential for future cancer treatment although further investigation in more animals and different organs are warranted.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.pan.2023.11.013DOI Listing

Publication Analysis

Top Keywords

novel treatment
8
treatment system
8
endoscopic ultrasound-guided
8
high-intensity focused
8
focused ultrasound
8
hifu transducer
8
hifu
7
novel
4
system endoscopic
4
ultrasound-guided high-intensity
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!