Measurement Accuracy When Using Spot Vision Screener With or Without Cycloplegia in Young Adults.

Clin Ophthalmol

Field of Orthoptics and Visual Sciences, Major in Medical and Rehabilitation Sciences, Niigata University of Health and Welfare Graduate School, Niigata, Japan.

Published: November 2023

Purpose: There are many unclear points about the accuracy of measurement of cycloplegic refraction using the Spot Vision Screener (SVS). This study aimed to investigate the accuracy of SVS measurements with cycloplegia for myopia.

Materials And Methods: Forty-nine healthy subjects were included, and refraction was measured. Objective refractions were measured by SVS, table-mounted autorefractometer (RT7000), and handheld autorefractometer (Retinomax Screeen) at noncycloplegic and cycloplegic conditions by 1% cyclopentolate. Subjective noncycloplegic refraction was obtained by a visual acuity and refraction test performed by certified orthoptists using a cross-cylinder. One-way repeated-measures analysis of variance was used to examine whether the measured refractions fluctuate due to different reflection tests.

Results: In the noncycloplegic condition, the mean (±standard deviation) spherical equivalent (SE) measured by subjective method, SVS, RT7000, and Retinomax Screeen were -2.56 ± 3.00, -2.62 ± 2.38, -3.05 ± 2.84, and -3.26 ± 2.97, respectively. The subjective SE and objective SE measured by SVS had significantly less myopic value than the objective SE measured by two autorefractometers (p < 0.001). In the cycloplegic condition, the mean (± standard deviation) SE measured by SVS, RT7000, and Retinomax Screeen were -2.07 ± 2.66, -2.62 ± 2.98, and -2.66 ± 3.02, respectively. The objective SE measured by SVS had significantly less myopic value than SEs measured using other methods (p < 0.001). In the cycloplegic condition, SVS showed a fixed error wherein the SE was more hyperopic than that with the subjective method and SVS had a proportional error.

Conclusion: In the measurement under cycloplegic conditions, use of an autorefractometer rather than a photorefractometer such as SVS was preferable.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10676106PMC
http://dx.doi.org/10.2147/OPTH.S431202DOI Listing

Publication Analysis

Top Keywords

measured svs
16
retinomax screeen
12
objective measured
12
svs
10
measured
9
spot vision
8
vision screener
8
measurement cycloplegic
8
cycloplegic conditions
8
subjective method
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!