Background: The 3D T1W turbo field echo sequence is a standard imaging method for acquiring high-contrast images of the brain. However, the contrast-to-noise ratio (CNR) can be affected by the turbo factor, which could affect the delineation and segmentation of various structures in the brain and may consequently lead to misdiagnosis. This study is aimed at evaluating the effect of the turbo factor on image quality and volumetric measurement reproducibility in brain magnetic resonance imaging (MRI).

Methods: Brain images of five healthy volunteers with no history of neurological diseases were acquired on a 1.5 T MRI scanner with varying turbo factors of 50, 100, 150, 200, and 225. The images were processed and analyzed with FreeSurfer. The influence of the TFE factor on image quality and reproducibility of brain volume measurements was investigated. Image quality metrics assessed included the signal-to-noise ratio (SNR) of white matter (WM), CNR between gray matter/white matter (GM/WM) and gray matter/cerebrospinal fluid (GM/CSF), and Euler number (EN). Moreover, structural brain volume measurements of WM, GM, and CSF were conducted.

Results: Turbo factor 200 produced the best SNR (median = 17.01) and GM/WM CNR (median = 2.29), but turbo factor 100 offered the most reproducible SNR (IQR = 2.72) and GM/WM CNR (IQR = 0.14). Turbo factor 50 had the worst and the least reproducible SNR, whereas turbo factor 225 had the worst and the least reproducible GM/WM CNR. Turbo factor 200 again had the best GM/CSF CNR but offered the least reproducible GM/CSF CNR. Turbo factor 225 had the best performance on EN (-21), while turbo factor 200 was next to the most reproducible turbo factor on EN (11). The results showed that turbo factor 200 had the least data acquisition time, in addition to superior performance on SNR, GM/WM CNR, GM/CSF CNR, and good reproducibility characteristics on EN. Both image quality metrics and volumetric measurements did not vary significantly ( > 0.05) with the range of turbo factors used in the study by one-way ANOVA analysis.

Conclusion: Since no significant differences were observed in the performance of the turbo factors in terms of image quality and volume of brain structure, turbo factor 200 with a 74% acquisition time reduction was found to be optimal for brain MR imaging at 1.5 T.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10665095PMC
http://dx.doi.org/10.1155/2023/6304219DOI Listing

Publication Analysis

Top Keywords

turbo factor
52
image quality
24
factor 200
20
turbo
17
gm/wm cnr
16
factor
14
factor image
12
cnr turbo
12
turbo factors
12
gm/csf cnr
12

Similar Publications

Cine-magnetic resonance imaging (MRI) has been used to track respiratory-induced motion of the liver and tumor and assist in the accurate delineation of tumor volume. Recent developments in compressed sensitivity encoding (SENSE; CS) have accelerated temporal resolution while maintaining contrast resolution. This study aimed to develop and assess hepatobiliary phase (HBP) cine-MRI scans using CS.

View Article and Find Full Text PDF

Comparing Large Language Model and Human Reader Accuracy with Image Challenge Case Image Inputs.

Radiology

December 2024

From the Department of Radiology, Research Institute of Radiological Science and Center for Clinical Imaging Data Science, Yonsei University College of Medicine, Seoul, Republic of Korea (P.S.S.); Department of Radiology and Research Institute of Radiology (W.H.S., C.H.S., K.J.P., P.H.K., S.J.C., Y.A., S.P., H.Y.P., N.E.O.), Department of Medical Science, Asan Medical Institute of Convergence Science and Technology (W.H.S., H.H.), and Department of Internal Medicine (C.Y.W.), Asan Medical Center, University of Ulsan College of Medicine, Olympic-ro 33, Songpa-gu, 05505 Seoul, Republic of Korea; University of Ulsan College of Medicine, Seoul, Republic of Korea (M.W.H.); Department of Orthopaedic Surgery, Seoul Seonam Hospital, Republic of Korea (S.T.C.); and Department of Pulmonary and Critical Care Medicine, Gumdan Top Hospital, Incheon, Republic of Korea (H.P.).

Background Application of multimodal large language models (LLMs) with both textual and visual capabilities has been steadily increasing, but their ability to interpret radiologic images is still doubted. Purpose To evaluate the accuracy of LLMs and compare it with that of human readers with varying levels of experience and to assess the factors affecting LLM accuracy in answering Image Challenge cases. Materials and Methods Radiologic images of cases from October 13, 2005, to April 18, 2024, were retrospectively reviewed.

View Article and Find Full Text PDF

Background: Previous studies explored MRI-based radiomic features for differentiating between human epidermal growth factor receptor 2 (HER2)-zero, HER2-low, and HER2-positive breast cancer, but deep learning's effectiveness is uncertain.

Purpose: This study aims to develop and validate a deep learning system using dynamic contrast-enhanced MRI (DCE-MRI) for automated tumor segmentation and classification of HER2-zero, HER2-low, and HER2-positive statuses.

Study Type: Retrospective.

View Article and Find Full Text PDF

Background: The long-term prevalence of clinically significant prostate cancer (csPCa) in patients with initial negative prostate biopsy is unknown.

Purpose: To investigate the rate of csPCa of men with initial negative biopsy.

Study Type: Retrospective analysis of prospectively collected data.

View Article and Find Full Text PDF

Background/objectives: This study aimed to investigate the accuracy of Tumor, Node, Metastasis (TNM) classification based on radiology reports using GPT3.5-turbo (GPT3.5) and the utility of multilingual large language models (LLMs) in both Japanese and English.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!