Measurement errors play an important role in the development of goniometric equipment, devices used to measure range of motion. Reasonable validity and reliability are critical for both the device and examiner before and after to testing in human subjects. The objective is to evaluate the concurrent validity and reliability of five different clinical goniometric devices for the purpose of establishing an acceptable measurement error margin for a novel device. We explored the validity and inter- and intrarater reliability scores of five goniometric devices namely (i) the universal goniometer (UG), a two-armed hand-held goniometer, (ii) the inclinometer (IC), featuring a single base, fluid level, and gravity-weighted inclinometer, (iii) the digital inclinometer (DI), functioning as both a DI and dynamometer, (iv) the smartphone application (SA), employing gyroscope-based technology within a smartphone platform application and (v) the modified inclinometer (MI), a gravity pendulum-based inclinometer equipped with a specialized fixing apparatus. Measurements were obtained at 12 standard angles and 8 human shoulder flexion angles ranging from 0° to 180°. Over two testing sessions, 120 standardized angle measurements and 160 shoulder angle measurements from 20 shoulders were repetitively taken by three examiners for each device. The intraclass correlation coefficient (ICC), standard error of measurement (SEM), and minimal detectable change (MDC) were calculated to assess reliability and validity. Concurrent validity was also evaluated through the execution of the 95% limit of agreement (95% LOA) and Bland-Altman plots, with comparisons made to the UG. The concurrent validity for all device pairs was excellent in both study phases (ICC > 0.99, 95% LOA - 4.11° to 4.04° for standard angles, and - 10.98° to 11.36° for human joint angles). Inter- and intrarater reliability scores for standard angles were excellent across all devices (ICC > 0.98, SEM 0.59°-1.75°, MDC 1°-4°), with DI showing superior reliability. For human joint angles, device reliability ranged from moderate to excellent (ICC 0.697-0.975, SEM 1.93°-4.64°, MDC 5°-11° for inter-rater reliability; ICC 0.660-0.996, SEM 0.77°-4.06°, MDC 2°-9° for intra-rater reliability), with SA demonstrating superior reliability. Wider angle measurement however resulted in reduced device reliability. In conclusion, our study demonstrates that it is essential to assess measurement errors independently for standard and human joint angles. The DI is the preferred reference for standard angle testing, while the SA is recommended for human joint angle testing. Separate evaluations across the complete 0°-180° range offer valuable insights.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10684565PMC
http://dx.doi.org/10.1038/s41598-023-48344-6DOI Listing

Publication Analysis

Top Keywords

concurrent validity
16
human joint
16
reliability
12
validity reliability
12
goniometric devices
12
standard angles
12
joint angles
12
clinical goniometric
8
measurement errors
8
inter- intrarater
8

Similar Publications

Purpose: Intestinal ischemia-reperfusion injury (IIRI) occurs as a result of temporary blood flow interruption, leading to tissue damage upon reperfusion. Oxidative stress plays a critical role in this process, instigating inflammation and cell death. Identifying and characterizing genes associated with the oxidative stress response can offer valuable insights into potential therapeutic targets for managing IIRI.

View Article and Find Full Text PDF

Background: Central nervous system tumors are among the most lethal types of cancer. A critical factor for tailored neurosurgical resection strategies depends on specific tumor types. However, it is uncommon to have a preoperative tumor diagnosis, and intraoperative morphology-based diagnosis remains challenging.

View Article and Find Full Text PDF

Background: Cognitive deterioration is common in multiple sclerosis (MS) and requires regular follow-up. Currently, cognitive status is measured in clinical practice using paper-and-pencil tests, which are both time-consuming and costly. Remote monitoring of cognitive status could offer a solution because previous studies on telemedicine tools have proved its feasibility and acceptance among people with MS.

View Article and Find Full Text PDF

Background: Telehomecare monitoring (TM) in patients with cancer is a complex intervention. Research shows variations in the benefits and challenges TM brings to equitable access to care, the therapeutic relationship, self-management, and practice transformation. Further investigation into these variations factors will improve implementation processes and produce effective outcomes.

View Article and Find Full Text PDF

Objective: To assess the psychometric properties of the Thai version of the Self-Care Self-Efficacy Scale version 3.0 (SCSES-v3.0) in individuals with chronic illnesses.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!