Purpose: Endovascular aortic aneurysm repair (EVAR) is the dominant treatment modality over open repair for abdominal aortic aneurysms. However, a higher rate of reinterventions remains the Achilles heel of EVAR. Although type 1A endoleak from proximal seal zone failure of EVAR remains one of the leading causes for reintervention, fenestrated branched devices suitable for proximal extension of failed EVAR are not widely available in the United States. Gore Thoracoabdominal Multibranch Endoprosthesis (TAMBE) is an off-the-shelf investigational device that provides supraceliac seal by incorporating 4 visceral and renal arteries via preloaded inner branches.

Case Report: In this article, we describe 2 cases of type 1A endoleak from previous EVAR devices repaired using TAMBE. Both cases were performed under the Food and Drug Administration (FDA) compassionate use exemption. Considerations on the case planning and implantation techniques of TAMBE specific to previous EVAR devices are reviewed.

Conclusions: Gore TAMBE can be utilized to repair a type 1A endoleak of a previous infrarenal EVAR device. Greater supraceliac coverage necessary for TAMBE relative to the minimal seal zone should be considered when applying this device for a type 1A endoleak.

Clinical Impact: This report demonstrates the feasibility of applying off-the-shelf TAMBE device to treat one of the most common failure modes of EVAR, type1A endoleak.

Download full-text PDF

Source
http://dx.doi.org/10.1177/15266028231214211DOI Listing

Publication Analysis

Top Keywords

type endoleak
16
gore thoracoabdominal
8
thoracoabdominal multibranch
8
multibranch endoprosthesis
8
endoprosthesis tambe
8
endovascular aortic
8
evar
8
seal zone
8
endoleak previous
8
previous evar
8

Similar Publications

Purpose: To report a case series on using a novel semi-branch feature in custom-made stent-grafts in the endovascular treatment of complex aortic aneurysms and summarize the contemporary usage of this technology.

Case Series: Four patients underwent endovascular aortic aneurysm repair (EVAR) with a custom-made semi-branch stent-graft (Semi-Branch Endovascular Aortic Aneurysm Repair [SBEVAR]). Two male patients, 75- and 76-year-old, were treated due to failed EVAR with late-type Ia endoleak, and the other two, 80- and 55-year-old male patients, due to a juxta-renal aortic abdominal aneurysm (JRAAA).

View Article and Find Full Text PDF

Objective: As aneurysmal disease is progressive, proximal disease progression and para-anastomotic aneurysms are complications experienced after open infrarenal abdominal aortic aneurysm repair (AAA). As such, fenestrated or branched endovascular repair (F/BEVAR) may be indicated in these patients. Data describing fenestrated endovascular aneurysm repair after prior open repair are limited to institutional databases.

View Article and Find Full Text PDF

Objective: The aim of this single-center case series is to demonstrate that an ultra-low dose (ULD) can be routinely achieved in the hybrid operating room in standard endovascular aortic repair (EVAR) for infrarenal abdominal aortic aneurysm by adjusting the manufacturer's predefined imaging parameters, hardware configurations and user protocols (including benchmarking).

Methods: The hybrid operating room manufacturer predefined EVAR software setup of the dose exposure control software (OPTIQ, Siemens Healthineers, Forchheim, Germany) at our university medical center was screened for possible improvements regarding radiation dose application. Tests on a water-equivalent as well as polymethyl methacrylate phantom model to assess the impact of technical settings were performed, including comparison of settings for exposure control software, different magnification, collimation configurations and detector distance.

View Article and Find Full Text PDF

We present a case of an 80-year-old female who presented with chest pain, vomiting and night sweats a few weeks post thoracic endovascular aortic aneurysm repair (TEVAR). A computed tomography (CT) scan demonstrated a type 1B endoleak for which she underwent a repeat TEVAR. Postoperatively, she developed fever, dysphagia, haematemesis and melaena.

View Article and Find Full Text PDF

Objective: We present our experience with endovascular Bentall procedure (Endo-Bentall) using a modular valve conduit (Endo-Bentall) in high-risk patients with aortic root pathologies.

Methods: The physician constructed Endo-Bentall device is composed of a self-expanding transcatheter aortic valve (TAVR), aortic endovascular stent graft (TEVAR), and two wire-reinforced fenestrations for coronary artery stenting. The TAVR valve is sutured into an appropriately sized TEVAR graft.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!