Purpose: Respiratory rate is a commonly used vital sign with various clinical applications. It serves as a crucial marker of acute health issues and any significant alteration in respiratory rate may be an early warning sign of major issues such as infections in the respiratory tract, respiratory failure, or cardiac arrest. Timely recognition of changes in respiratory rate enables prompt medical action, while neglecting to detect a change may lead to adverse patient outcomes. Here, we report on the performance of respiratory rate determined using a depth sensing camera system (RR) which allows for continuous, non-contact 'touchless' monitoring of this important vital sign.

Methods: Thirty adult volunteers undertook a range of set breathing rates to cover a target breathing range of 4-40 breaths/min. Depth information was acquired from the torso region of the subjects using an Intel D415 RealSense camera positioned above the bed. The depth information was processed to generate a respiratory signal from which RR was calculated. This was compared to a manually scored capnograph reference (RR).

Results: An overall RMSD accuracy of 0.77 breaths/min was achieved across the target respiratory rate range with a corresponding bias of 0.05 breaths/min. This corresponded to a line of best fit given by RR = 1.01 x RR - 0.22 breaths/min with an associated high degree of correlation (R = 0.997). A breakdown of the performance with respect to sub-ranges corresponding to respiratory rates or ≤7, >7-10, >10-20, >20-30, >30 breaths/min all exhibited RMSD accuracies of less than 1.00 breaths/min. We also had the opportunity to test the performance of spontaneous breathing of the subjects which occurred during the study and found an overall RMSD accuracy of 1.20 breaths/min with corresponding accuracies ≤1.30 breaths/min across each of the individual sub-ranges.

Conclusions: We have conducted an investigative study of a prototype depth sensing camera system for the non-contact monitoring of respiratory rate. The system achieved good performance with high accuracy across a wide range of rates including both clinically important high and low rates.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.rmed.2023.107463DOI Listing

Publication Analysis

Top Keywords

respiratory rate
28
respiratory
11
depth sensing
8
sensing camera
8
camera system
8
breaths/min
8
rmsd accuracy
8
rate
7
accurate continuous
4
continuous respiratory
4

Similar Publications

Background: Type 1 diabetes is the most common endocrine health condition among youth. Healthcare professionals must consider evidence-based guidelines in managing children and adolescents with diabetic ketoacidosis (DKA). The current study aims to assess the outcomes of implementing clinical guidelines by the American Diabetes Association to manage DKA among pediatrics in an emergency department in Palestine.

View Article and Find Full Text PDF

Complex biological systems undergo sudden transitions in their state, which are often preceded by a critical slowing down of dynamics. This results in longer recovery times as systems approach transitions, quantified as an increase in measures such as the autocorrelation and variance. In this study, we analysed paediatric patients in intensive care for whom mechanical ventilation was discontinued through removal of the endotracheal tube (extubation).

View Article and Find Full Text PDF

Objective: The ADVOCATE trial demonstrated that treatment of active granulomatosis with polyangiitis (GPA) and microscopic polyangiitis (MPA) with avacopan was noninferior in achieving remission at week 26 and superior for sustained remission at week 52 compared with a prednisone taper. This analysis of ADVOCATE evaluated the efficacy and safety of avacopan in patients with ear, nose, throat (ENT), or lung manifestations.

Methods: This post hoc analysis included patients enrolled in ADVOCATE with ENT or lung manifestations at baseline.

View Article and Find Full Text PDF

Background: Multiplex genetic testing is recommended when treating nonsmall cell lung cancer. A certain percentage of test failures in RNA assays owing to poor surgical specimen quality have been documented, and fixation failure is a possible cause. At our institution, sheet-like fixation is performed to reduce fixation time.

View Article and Find Full Text PDF

A20 as a Potential Therapeutic Target for COVID-19.

Immun Inflamm Dis

January 2025

State Key Laboratory of Oral Diseases and National Clinical Research Center for Oral Diseases, West China Hospital of Stomatology, Sichuan University, Chengdu, China.

Background: Coronavirus disease 2019 (COVID-19), caused by the severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2), is a major concern due to its astonishing prevalence and high fatality rate, especially among elderly people. Patients suffering from COVID-19 may exhibit immunosuppression in the initial stage of infection, while a cytokine storm can occur when the disease progresses to a severe stage. This inopportune immune rhythm not only makes patients more susceptible to the virus but also leads to numerous complications resulting from the excessive production of inflammatory factors.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!