Background: Coronary accessibility following redo-transcatheter aortic valve replacement (redo-TAVR) is increasingly important, particularly in younger low-risk patients. This study aimed to predict coronary accessibility after simulated Sapien-3 balloon-expandable valve implantation within an Evolut supra-annular, self-expanding valve using pre-TAVR computed tomography (CT) imaging.

Methods: A total of 219 pre-TAVR CT scans from the Evolut Low-Risk CT substudy were analyzed. Virtual Evolut and Sapien-3 valves were sized using CT-based diameters. Two initial Evolut implant depths were analyzed, 3 and 5 mm. Coronary accessibility was evaluated for 2 Sapien-3 in Evolut implant positions: Sapien-3 outflow at Evolut node 4 and Evolut node 5.

Results: With a 3-mm initial Evolut implant depth, suitable coronary access was predicted in 84% of patients with the Sapien-3 outflow at Evolut node 4, and in 31% of cases with the Sapien-3 outflow at Evolut node 5 (<0.001). Coronary accessibility improved with a 5-mm Evolut implant depth: 97% at node 4 and 65% at node 5 (<0.001). When comparing 3- to 5-mm Evolut implant depth, sinus sequestration was the lowest with Sapien-3 outflow at Evolut node 4 (13% versus 2%; <0.001), and the highest at Evolut node 5 (61% versus 32%; <0.001).

Conclusions: Coronary accessibility after Sapien-3 in Evolut redo-TAVR relates to the initial Evolut implant depth, the Sapien-3 outflow position within the Evolut, and the native annular anatomy. This CT-based quantitative analysis may provide useful information to inform and refine individualized preprocedural CT planning of the initial TAVR and guide lifetime management for future coronary access after redo-TAVR.

Registration: URL: https://www.clinicaltrials.gov; Unique identifier: NCT02701283.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10653288PMC
http://dx.doi.org/10.1161/CIRCINTERVENTIONS.123.013238DOI Listing

Publication Analysis

Top Keywords

evolut node
16
coronary accessibility
12
evolut implant
12
sapien-3 outflow
12
outflow evolut
12
evolut
11
coronary access
8
computed tomography
8
initial evolut
8
sapien-3
6

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!