Influence of ionization volume and sample gas flow rate on separation power in gas chromatography-ion mobility spectrometry.

J Chromatogr A

Department of Sensors and Measurement Technology, Institute of Electrical Engineering and Measurement Technology, Leibniz University Hannover, Appelstr. 9A, Hannover 30167, Germany.

Published: January 2024

In this work, the influence of the sample gas flow rate and the ionization region volume of an ion mobility spectrometer (IMS) used as a detector in gas chromatography (GC) on GC-IMS peak shape has been investigated. Therefore, a drift tube IMS with a field-switching ion shutter, a defined ionization region volume and an ultra-violet radiation source was used. To identify the influence of the sample gas flow rate entering the ionization region (equals the GC carrier gas flow rate if no further make-up gas is used) and the ionization region volume on peak broadening and signal intensity, different sample volumes as they would elute from a GC were tested at a variety of sample gas flow rates at a given ionization region volume. The results clearly show that for low sample gas flow rates a depletion of sample molecules in the ionization region leads to a significant decrease in effective detector volume but also to reduced signal intensities. Therefore, for optimal performance of a GC-IMS, the optimal operating point of the GC should match the flow range, where the IMS provides the best compromise between signal-to-noise ratio and peak broadening.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.chroma.2023.464506DOI Listing

Publication Analysis

Top Keywords

gas flow
24
ionization region
24
sample gas
20
flow rate
16
region volume
16
gas
9
influence sample
8
peak broadening
8
flow rates
8
sample
7

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!