Goals: We aimed to determine the performance of the OC-Auto Micro 80 fecal immunochemical test (FIT) in an average-risk population receiving care in an integrated, academic-community health system.

Background: The FIT is the most used colorectal cancer (CRC) screening test worldwide. However, many Food and Drug Administration-cleared FIT products have not been evaluated in clinical settings.

Study: We performed a retrospective cohort study of patients (50 to 75 y old) in the University of Washington Medicine health care system who were screened for CRC by OC-Auto Micro 80 FIT between March 2016 and September 2021. We used electronic health records to extract patient-level and clinic-level factors, FIT use, colonoscopy, and pathology findings. The primary outcomes were the FIT positivity rate and neoplasms detected at colonoscopy. Secondary outcomes were FIT positivity by sex and safety-net versus non-safety-net clinical settings.

Results: We identified 39,984 FITs completed by 26,384 patients; 2411 (6.0%) had a positive FIT result (>100 ng/mL of hemoglobin in buffer), and 1246 (51.7%) completed a follow-up colonoscopy. The FIT positive rate was 7.0% in men and 5.2% in women ( P <0.01). Among those who completed a colonoscopy after an abnormal FIT result, the positive predictive value for CRC, advanced adenoma, and advanced neoplasia was 3.0%, 20.9%, and 23.9%, respectively.

Conclusions: In a retrospective analysis of a large heterogeneous population, the OC-Auto Micro 80 FIT for CRC screening demonstrated a positivity rate of 6.0% and a positive predictive value for CRC of 3.0%.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10963337PMC
http://dx.doi.org/10.1097/MCG.0000000000001928DOI Listing

Publication Analysis

Top Keywords

oc-auto micro
12
fit
9
performance oc-auto
8
micro fecal
8
fecal immunochemical
8
immunochemical test
8
integrated academic-community
8
academic-community health
8
outcomes fit
8
fit positivity
8

Similar Publications

Goals: We aimed to determine the performance of the OC-Auto Micro 80 fecal immunochemical test (FIT) in an average-risk population receiving care in an integrated, academic-community health system.

Background: The FIT is the most used colorectal cancer (CRC) screening test worldwide. However, many Food and Drug Administration-cleared FIT products have not been evaluated in clinical settings.

View Article and Find Full Text PDF

Background: Rectal bleeding is the most important symptom of intestinal neoplasia; thus, tests of occult blood detection in stools are widely used for pre neoplastic lesions and colorectal cancer (CRC) screening.

Objective: Evaluate the accuracy of OC-Sensor quantitative test (Eiken Chemical, Tokyo, Japan) at cut-off 10 µg Hb/g feces (50 ng/mL) in a cohort of subjects that had to undergo diagnostic colonoscopy, and if more than one sample collected in consecutive days would improve the diagnostic accuracy of the test.

Methods: Patients (mean age 56.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!