We report the design and experimental validation of a compact positron emission tomography (PET) detector module (DM) intended for building a preclinical PET and electron-paramagnetic-resonance-imaging hybrid system that supports sub-millimeter image resolution and high-sensitivity, whole-body animal imaging. The DM is eight detector units (DU) in a row. Each DU contains 12×12 lutetium-yttrium oxyorthosilicate (LYSO) crystals having a 1.05 mm pitch read by 4×4 silicon photomultipliers (SiPM) having a 3.2 mm pitch. A small-footprint, highly-multiplexing readout employing only passive electronics is devised to produce six outputs for the DM, including two outputs derived from SiPM cathodes for determining event time and active DU and four outputs derived from SiPM anodes for determining energy and active crystal. Presently, we have developed two DMs that are 1.28×10.24 cm in extent and approximately 1.8 cm in thickness, with their outputs sampled at 0.7 GS/s and analyzed offline. For both DMs, our results show successfully discriminated DUs and crystals. With no correction for SiPM nonlinearity, the average energy resolution for crystals in a DU ranges from 14% to 16%. While not needed for preclinical imaging, the DM may support 300-400 ps time-of-flight resolution.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10655702PMC
http://dx.doi.org/10.1109/trpms.2023.3301788DOI Listing

Publication Analysis

Top Keywords

positron emission
8
emission tomography
8
tomography pet
8
pet electron-paramagnetic-resonance-imaging
8
hybrid system
8
pet detector
8
detector module
8
outputs derived
8
derived sipm
8
preclinical positron
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!