Distance-Based Self-Powered Signal Transduction of Ion-Selective Electrodes to an Electronic Paper Display Array.

Anal Chem

Department of Inorganic and Analytical Chemistry, University of Geneva, Quai Ernest-Ansermet 30, CH-1211 Geneva, Switzerland.

Published: December 2023

In this article, we report on the first distance-based readout self-powered potentiometric sensor. The approach is considered more user-friendly for detection by the naked eye and is less prone to optical interferences compared with a direct observation of the pixel darkening. pH-selective electrodes were chosen as a model system to demonstrate the principle in which seven bar-shaped pixels connected in series on one e-paper share one common ground. By connecting each of the pixels serially to capacitors of different capacitances, the fraction of the measurement cell voltage loaded onto the pixels becomes controllable. Consequently, the pixels give different gray values when powered by the same ion-selective electrode (ISE). As a result, the pH information on the sample is visualized as a distance-based signal and the dependence between the capacitance and 1/ (the reciprocal slope in the relationship between absorbance and pH) was constructed. In the current system, a 1 μF capacitance difference changes the value of 1/ by 4.18. With the current setup, the pH accuracy is about 0.5 when comparing the e-paper output to a color card.

Download full-text PDF

Source
http://dx.doi.org/10.1021/acs.analchem.3c03994DOI Listing

Publication Analysis

Top Keywords

distance-based self-powered
4
self-powered signal
4
signal transduction
4
transduction ion-selective
4
ion-selective electrodes
4
electrodes electronic
4
electronic paper
4
paper display
4
display array
4
array article
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!