Objectives: FebriDx is a CE-marked, FDA-approved point-of-care test that detects the antiviral host response protein Myxovirus Resistance Protein A (MxA), in addition to C-reactive protein, using finger-prick blood. FebriDx MxA detection had a high negative predictive value for COVID-19 in symptomatic adults presenting to hospital in the first waves of the pandemic and was used subsequently as a 'rule out' triage tool in Emergency departments. The diagnostic accuracy of FebriDx MxA in the current context of co-circulation of influenza, SARS-CoV-2, and Respiratory Syncytial Virus (RSV), and in the era of COVID-19 vaccination, is unknown.
Methods: We retrospectively evaluated the diagnostic performance of FebriDx MxA in adults with acute respiratory symptoms presenting to the Emergency Department (ED) of a large UK teaching hospital using Reverse Transcription Polymerase Chain Reaction (RT-PCR) as the reference standard (Cepheid Xpert Xpress SARS-CoV-2/Flu/RSV).
Results: Between March 9th 2022 and March 8th 2023, 5426 patients had both FebriDx and RT-PCR testing with valid results. 999 (18.4%) of patients had influenza detected, 520 (9.6%) SARS-CoV-2, and 190 (3.5%) RSV. Negative Predictive Value (NPV) of MxA detection by FebriDx was 97.5% (96.9-98.0) for influenza, 97.1% (96.4-97.7) for SARS-CoV-2, 98.1% (97.5-98.6) for RSV, and 92.8% (91.8-93.7) for all viruses combined.
Conclusions: In symptomatic adults, FebriDx MxA had a high NPV for influenza and RSV, and retained a high NPV for SARS-CoV-2, in the context of virus co-circulation and widespread COVID-19 vaccination. FebriDx continues to be a useful 'rule out' triage tool in the ED and could potentially be scaled to provide a national triage solution for future viral pandemics.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.jinf.2023.11.003 | DOI Listing |
J Antimicrob Chemother
June 2024
Primary Care Research Centre, Aldermoor Health Centre, Aldermoor Close, University of Southampton, Southampton SO16 5ST, UK.
Introduction: FebriDx® is a CE-marked, single-use point-of-care test with markers for bacterial [C-reactive protein (CRP)] and viral [myxovirus resistance protein A (MxA)] infection, using finger-prick blood samples. Results are available after 10-12 min. We explored the usability and potential impact of FebriDx® in reducing antibiotic prescriptions for lower respiratory tract infection (LRTI) in primary care, and the feasibility of conducting a randomized controlled trial (RCT).
View Article and Find Full Text PDFBJGP Open
October 2024
Primary Care Research Centre, University of Southampton, Southampton, UK.
J Clin Med
December 2023
Department of Emergency Medicine, Erasmus University Medical Center, 3015 GD Rotterdam, The Netherlands.
FebriDx is a rapid point-of-care test combining qualitative measurements of C-reactive protein (CRP) and Myxovirus Resistance Protein A (MxA) using a disposable test device to detect and differentiate acute bacterial from viral respiratory tract infections. The goal of this study was to investigate the diagnostic accuracy of FebriDx in patients with suspected respiratory tract infections in the emergency department (ED). This was an observational cohort study, performed in the ED of an academic hospital.
View Article and Find Full Text PDFJ Infect
January 2024
School of Clinical and Experimental Sciences, Faculty of Medicine, University of Southampton, Southampton, UK; Department of Infection, University Hospital Southampton NHS Foundation Trust, Southampton, UK; NIHR Southampton Biomedical Research Centre, University Hospital Southampton NHS Foundation Trust, Southampton, UK.
Objectives: FebriDx is a CE-marked, FDA-approved point-of-care test that detects the antiviral host response protein Myxovirus Resistance Protein A (MxA), in addition to C-reactive protein, using finger-prick blood. FebriDx MxA detection had a high negative predictive value for COVID-19 in symptomatic adults presenting to hospital in the first waves of the pandemic and was used subsequently as a 'rule out' triage tool in Emergency departments. The diagnostic accuracy of FebriDx MxA in the current context of co-circulation of influenza, SARS-CoV-2, and Respiratory Syncytial Virus (RSV), and in the era of COVID-19 vaccination, is unknown.
View Article and Find Full Text PDFJAMA Netw Open
October 2022
Medical Affairs, Lumos Diagnostics, Sarasota, Florida.
Importance: Acute respiratory infections (ARIs) account for most outpatient visits. Discriminating bacterial vs viral etiology is a diagnostic challenge with therapeutic implications.
Objective: To investigate whether FebriDx, a rapid, point-of-care immunoassay, can differentiate bacterial- from viral-associated host immune response in ARI through measurement of myxovirus resistance protein A (MxA) and C-reactive protein (CRP) from finger-stick blood.