Objective: We assessed whether the immediate pressure adjustments provided by a phacoemulsification handpiece with a built-in pressure sensor (Active Sentry, AS) could decrease the amount of energy delivered in the eye during cataract surgery.
Methods And Analysis: The Study of Active Sentry in Cataract Surgery was a multicentre prospective clinical trial. The Centurion system (Alcon Labs) was used and we compared phacoemulsifications performed with the AS handpiece with procedures using a sensor placed in the phacoemulsifier (non-AS). The primary outcome measure was the cumulative dissipated energy (CDE) used during the procedures. The secondary outcome measures were the total, longitudinal and torsional ultrasound (US) times, the duration of the surgery and the amount of fluid used during the surgeries.
Results: The study analysed 1432 (800 AS and 632 non-AS) procedures. The mean patient age was 72.4±10.0 years. The median CDE was respectively 6.3 (IQR 3.5-9.9) and 6.7 (IQR 4.5-11.5) with and without AS (p=0.0001). The median US time was, respectively, 48.0 s (IQR 34.0-68.0) and 55.0 s (IQR 38.0-80.0) with and without AS (p<0.0001). Torsional and longitudinal US times were reduced with AS. The median duration of the procedure was, respectively, 9.0 min (IQR 7.0-12.0) and 10.0 min (IQR 7.0-13.0) (p=0.0002) with and without AS. The median balanced salt solution volume used was 52.0 cm (IQR 41.0-72.0) and 57.0 cm (IQR 42.0-81.0) with and without AS (p=0.0018).
Conclusion: The phacoemulsifications performed with a pressure sensor built in the handpiece delivered less energy in the eye and were shorter.
Trial Registration Number: NCT04732351.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10661016 | PMC |
http://dx.doi.org/10.1136/bmjophth-2023-001431 | DOI Listing |
BMJ Open Ophthalmol
November 2023
Université Paris Cité, URC Paris Centre, Paris, France.
Objective: We assessed whether the immediate pressure adjustments provided by a phacoemulsification handpiece with a built-in pressure sensor (Active Sentry, AS) could decrease the amount of energy delivered in the eye during cataract surgery.
Methods And Analysis: The Study of Active Sentry in Cataract Surgery was a multicentre prospective clinical trial. The Centurion system (Alcon Labs) was used and we compared phacoemulsifications performed with the AS handpiece with procedures using a sensor placed in the phacoemulsifier (non-AS).
J Endod
March 2022
Department of Restorative Dentistry, Faculty of Dentistry, Near East University, Nicosia, Turkey.
Introduction: To demonstrate the unknown kinematics of complex adaptive motion with respect to disclosed values, the aim of this study was (1) to analyze the adaptive motion and (2) to compare the effects of constant torque values on kinematics.
Methods: A new endodontic motor (Aseptico AEU-28SYB Elements; SybronEndo, Orange, CA) and 8:1 contra-angle handpiece were used for analysis. A pilot test was conducted for synchronizing the external experimental stress on the shaft of the target object and the built-in torque sensor of the motor source.
Ophthalmic Plast Reconstr Surg
October 2014
*Department of Ophthalmology, West Virginia University, Morgantown, West Virginia; †Department of Ophthalmology, Harvard Medical School; ‡Ophthalmic Plastic and Reconstructive Surgery, Massachusetts Eye and Ear Infirmary; §Department of Otology and Laryngology, Harvard Medical School; and ‖Department of Otolaryngology, Massachusetts Eye and Ear Infirmary, Boston, Massachusetts, U.S.A.
Purpose: Stereotactic navigation systems have been used in neurosurgery and otolaryngology with great success. The current investigation illustrates the novel use of a microdebrider with built-in stereotactic guidance in a series of thyroid orbitopathy patients who underwent deep lateral orbital wall decompression surgery.
Methods: A noncomparative, interventional, retrospective case series of patients who underwent deep lateral deep orbital wall decompression from 2006 to 2013 was conducted in accordance with Institutional Review Board policy and the Declaration of Helsinki.
Lasers Surg Med
February 2008
Division of Facial Plastic and Reconstructive Surgery, Department of Otology and Laryngology, Massachusetts Eye and Ear Infirmary, Harvard Medical School, Boston, Massachusetts, USA.
Objective: Various energy delivery systems have been utilized to treat superficial rhytids in the aging face. The Intense Ultrasound System (IUS) is a novel modality capable of transcutaneously delivering controlled thermal energy at various depths while sparing the overlying tissues. The purpose of this feasibility study was to evaluate the response of porcine tissues to various IUS energy source conditions.
View Article and Find Full Text PDFOrtholux XT is a high-intensity light source emitting filtered visible blue light in the 400- to 500-nanometer range for polymerization of visible-light-cured resins. The Ortholux handpiece comes with a portable power supply, a light-intensity check in the power supply, a spare lamp, an eyeshield and a mounting kit. The handpiece consists of a pistol grip with a thermoplastic housing that contains the light source, cooling fan, light guide receptacle (8- or 13-millimeter-diameter fused quartz lightguide), optical filter, light switch and timer switch.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!