Background: Surgical clips are commonly used during laparoscopic cholecystectomy for cystic duct and artery ligation. Titanium and polymer clips are the two most common types used for this indication. Given the cost-saving potential, design advantages, and decreased incidence of complications associated with polymer clips, we sought to study whether there is a clinically significant difference in outcome between polymer and titanium clips in laparoscopic cholecystectomy.

Methods: Fifty consecutive cases using polymer clips followed by 50 consecutive cases using metal clips over a 6-month period by residents under the direction of a single surgeon were retrospectively reviewed. The following outcomes were evaluated: incidence of bile leak, postoperative bleeding, need for additional procedures, hospital length of stay, and cost.

Results: We found that significantly more misfires occurred with the use of the polymer clips (n=17) than with the titanium clips (n=2, p<.001). Eight cases (16%) required opening of an additional polymer clip cartridge to complete the operation. Despite this additional expense, the total cost as it pertained to clip usage ($30.32 USD) was still lower than that using titanium clips ($139.17 USD). While these numbers were not statistically significant, three cases had bile leaks and required additional procedures, all of which were performed with metal clips. No postoperative bleeds were identified and there was no difference in hospital length of stay; most patients were discharged on the day of the procedure.

Conclusion: These findings demonstrate comparable clinical outcomes between laparoscopic cholecystectomies performed with polymer and titanium clips, though polymer clip usage carries a lower cost.

Download full-text PDF

Source
http://dx.doi.org/10.52198/23.STI.43.GS1730DOI Listing

Publication Analysis

Top Keywords

polymer clips
16
titanium clips
12
clips
9
clips laparoscopic
8
laparoscopic cholecystectomy
8
consecutive cases
8
polymer
6
polymer versus
4
titanium
4
versus titanium
4

Similar Publications

Laparoscopic cholecystectomy is a widely performed procedure, with securing the cystic duct being a critical step to prevent bile leakage. Traditional metal clips are commonly used, but alternative methods, such as non-absorbable polymer clips, absorbable clips, sutures, and ultrasonic shears, are also utilized. This systematic review and meta-analysis evaluates the safety and efficacy of various cystic duct securing techniques.

View Article and Find Full Text PDF

Introduction Lymphocele is a typical complication of pelvic lymph node dissection (PLND) in robot-assisted radical prostatectomy (RARP). This study aimed to compare postoperative lymphatic leakage between the polymer ligation clip and vessel sealer, and evaluated the costs associated with the former. Methods The study enrolled patients who underwent RARP with PLND at our institution between April 2018 and March 2023 and were treated with a vessel sealer (LigaSure Blunt Tip 44 NC; Medtronic, Dublin, Ireland) until September 2021, and Hem-o-lok polymer ligation clips(Teleflex, Wayne, PA, USA) thereafter.

View Article and Find Full Text PDF

During gastric surgery, the stomach wall is compressed with clamps and sutures or staple lines. These short- and long-term deformations can severely compromise the integrity of the tissue and make it difficult for the stomach wall to respond and remodel to the new loading conditions. Consequently, serious intra- and postoperative complications such as the formation of leaks during bariatric surgeries, can be associated with these immense tissue deformations.

View Article and Find Full Text PDF

Objective: To develop and validate a device that measures the pressure exerted by forceps on the fetal head for clinical use.

Background: The lack of clinical tools to quantify forceps pressure on the fetal head may impact maternal and neonatal outcomes. Existing studies have not measured the direct contact pressure between forceps blades and the fetal head, highlighting the need for innovation.

View Article and Find Full Text PDF

In addition to the basic and main parts of hospital equipment, 316 L stainless steel is widely utilized in futures such as nails and screws, wires and medical bone clips, dental implants, heart springs (stents), needles, surgical scissors, etc. In the present study, the electrophoretic deposition of a composite based on chitosan (CS), gelatin, nano and microparticles of hydroxyapatite on a 316 L stainless steel substrate was investigated. Hydroxyapatite particles are added to it due to the ossification abilities of steel and due to an enhanced adhesion and bone production, CS and biocompatible gelatin polymer particles were also added to hydroxyapatite.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!