Purpose: In total knee arthroplasty (TKA) with posterior cruciate ligament (PCL) retention, the medial and lateral insert conformity that restores in vivo native (i.e., healthy) knee tibial rotation and high function without causing stiffness is unknown. The purpose was to determine whether a ball-in-socket (B-in-S) medially conforming (MC) and flat lateral insert implanted with unrestricted kinematic alignment (KA) TKA and PCL retention restores tibial rotation to native.

Methods: One group of 25 patients underwent unrestricted KA TKA with manual instruments. Another group of 25 patients had native knees. Single-plane fluoroscopy imaged each knee while patients performed step-up and chair rise activities. Following 3D model-to-2D image registration, anterior-posterior (A-P) positions of the femoral condyles were determined. Changes in A-P positions with flexion were used to determine tibial rotation.

Results: At maximum flexion, mean tibial rotations of KA TKA knees were comparable to native knees (Step up: 12.3° ± 4.4° vs. 13.1° ± 12.0°, p = 0.783; Chair Rise: 12.7° ± 6.2° vs. 12.6° ± 9.5º, p = 0.941). However, paths of rotation differed in that screw home motion was less evident in KA TKA knees. At 8 months follow-up, the median Forgotten Joint Score was 69 points (range 65 to 85), the median Oxford Knee Score was 43 points (range 40 to 46), and mean knee flexion was 127º ± 8°.

Conclusions: The ball-in-socket medial, flat lateral insert and PCL retention implanted with unrestricted KA TKA restored in vivo native knee tibial rotation at maximum flexion for each activity and high function without stiffness. Providing high A-P stability, this implant design might benefit patients desiring to return to demanding work and recreational activities.

Level Of Evidence: Therapeutic - Level II.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10646131PMC
http://dx.doi.org/10.1186/s40634-023-00671-3DOI Listing

Publication Analysis

Top Keywords

tibial rotation
16
knee tibial
12
pcl retention
12
lateral insert
12
ball-in-socket medial
8
posterior cruciate
8
cruciate ligament
8
unrestricted kinematic
8
kinematic alignment
8
high function
8

Similar Publications

Effect of surgical treatment on relative tibial external rotation in patients with recurrent patellar dislocation.

Knee

December 2024

Graduate School of Guangzhou University of Chinese Medicine, Guangzhou, Guangdong, China; Second Affiliated Hospital of Guangzhou Medical University, Guangzhou, Guangdong, China. Electronic address:

Background: In the knee joints of patients with recurrent patellar dislocation (RPD), an increased relative tibial external rotation (rTER) is often observed. However, the changes in this parameter pre- and postoperatively remain unclear.

Purpose: The purpose of this study was to reveal preoperative and postoperative changes in rTER in patients with RPD.

View Article and Find Full Text PDF

Objective: Accurate rotational reduction following tibial shaft fracture fixation is absent in up to 36% of cases yet may be critical for lower extremity biomechanics. The objective of this cadaveric study was to compare the results of freehand methods of reduction with software-assisted reduction.

Methods: Four fellowship-trained orthopaedic trauma surgeons attempted rotational correction in a cadaveric model with fluoroscopic assistance (without radiographic visualization of the fracture site) using (1) their method of choice (MoC) and (2) software assistance (SA).

View Article and Find Full Text PDF

Purpose: To clarify the influence of biomechanics on post-operative clinical outcomes in bicruciate-retaining total knee arthroplasty (BCR-TKA).

Methods: Severe medial osteoarthritis who underwent BCR-TKA were examined. Each patient was asked to perform a squat (weight-bearing [WB]) and active assisted knee flexion (non-WB [NWB]) under single fluoroscopy surveillance.

View Article and Find Full Text PDF

Background: Long-leg alignment and joint line obliquity have traditionally been assessed using two-dimensional (2D) radiography, but the accuracy of this measurement has remained unclear. This study aimed to evaluate the accuracy of 2D measurements of lateral distal femoral angle (LDFA) and medial proximal tibial angle (MPTA) using upright three-dimensional (3D) computed tomography (CT).

Methods: This study involved 66 knees from 38 patients (34 women, four men) with knee osteoarthritis (OA), categorized by Kellgren-Lawrence (KL) grade.

View Article and Find Full Text PDF

Introduction: As the number of revision total knee arthroplasties (rTKA) continues to rise, there is increasing interest in the use of contemporary rotating hinge prostheses. These devices often incorporate porous cones to fill bone defects and enhance long-term fixation. This study evaluated the clinical and functional outcomes and survivorship in rTKA patients utilizing a rotating hinge prosthesis with flexible titanium (FT) cones, porous tantalum (PT) cones, or no cones.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!