Energy-based scatter estimation in clinical PET.

Med Phys

Department of Radiology, Perelman School of Medicine, University of Pennsylvania, Philadelphia, Pennsylvania, USA.

Published: January 2024

Background: Scatter correction (SC) is essential in PET for accurate quantitative imaging. The state-of-the-art SC method is single-scatter simulation (SSS). Although this method is usually robust and accurate, it can fail in some situations, for example when there is motion between the CT and PET scans in PET/CT. Therefore, it is of interest to consider other SC methods.

Purpose: In this work, an energy-based scatter estimation (EBS) method is described in detail, tested in phantoms and patients, and compared to SSS.

Methods: This version of EBS was developed for list-mode data from Biograph Vision-600 PET/CT scanner. EBS is based on digitized 2D energy histograms in each bin of a coarsely sampled PET sinogram, either with or without time of flight (TOF). The histograms are modeled as a noisy realization of a linear combination of nine basis functions whose parameters were derived from a measurement of the 511-keV photopeak spectrum as well as Monte-Carlo simulations of the scattering process. EBS uses an iterative expectation maximization approach to determine the coefficients in the linear combination, and from this estimates the scatter. The investigation was restricted to F-based PET data in which the acquired number of counts was similar to the levels seen in oncological whole-body PET/CT scans. To evaluate the performance, phantom scans were used that involved the NEMA NU2-2018 protocol, a slab phantom, an NU 2-1994 phantom, a cardiac phantom in an anthropomorphic chest phantom, and a uniformly-filled torso phantom with a bladder phantom slightly outside the axial field of view. Contrast recovery (CR) and other parameters were evaluated in images reconstructed with SSS and EBS. Furthermore, FDG PET scans of seven lung cancer patients were used in the evaluation. Standardized uptake values (SUV) based on SSS and EBS were compared in 27 lesions.

Results: EBS and SSS images were visually similar in all cases except the torso + bladder phantom, where the EBS was much closer to the expected uniform image. The NU2-2018 analysis indicated a 2% scatter residual in EBS images compared to 3% with SSS, and 10% higher background variability, which is a surrogate for image noise. The cardiac phantom scan showed that CR was 98.2% with EBS and 99.6% with SSS, and that the SSS sinogram had values greater than the net-true emission sinogram, indicating a slight overcorrection in the case of SSS. In the lesion SUV comparison in patient scans, EBS correlated strongly (R  = 0.9973) with SSS, and SUV based on EBS were systematically 0.1 SUV lower. In the case of the torso + bladder phantom portion, the SSS image of the torso + bladder phantom was 299% times hotter than expected in one area, due to scatter estimation error, compared to 16% colder with EBS.

Conclusions: In evaluating clinically relevant parameters such as SUV in focal lesions, EBS and SSS give almost the same results. In phantoms, some scatter figures of merit were slightly improved by use of EBS, though an image variability figure of merit was slightly degraded. In typical oncological whole-body PET/CT, EBS may be a suitable replacement for SSS, especially when SSS fails due to technical problems during the scan.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC11483182PMC
http://dx.doi.org/10.1002/mp.16826DOI Listing

Publication Analysis

Top Keywords

bladder phantom
16
ebs
15
sss
13
scatter estimation
12
torso bladder
12
phantom
11
energy-based scatter
8
pet scans
8
linear combination
8
oncological whole-body
8

Similar Publications

Radiation dose and diagnostic image quality are opposing constraints in x-ray CT. Conventional methods do not fully account for organ-level radiation dose and noise when considering radiation risk and clinical task. In this work, we develop a pipeline to generate individualized organ-specific dose and noise at desired dose levels from clinical CT scans.

View Article and Find Full Text PDF

Antimicrobial resistance (AMR) poses a significant global threat to public health systems, rendering antibiotics ineffective in treating infectious diseases. Combined use of bio compounds, including bacteriophages and plant extracts, is an attractive approach to controlling antibiotic resistance. In this study, the combination of phage cocktail (Isf-Pm1 and Isf-Pm2) and crude extract (AME) was investigated in controlling biofilm-forming multi-drug resistant isolates, and a phantom bladder model.

View Article and Find Full Text PDF

Background: Huntington's disease (HD) is a rare neurodegenerative disorder caused by an expansion of the CAG trinucleotide repeat in the huntingtin gene which encodes the mutant huntingtin protein (mHTT) that is associated with HD-related neuropathophysiology. Noninvasive visualization of mHTT aggregates in the brain, with positron emission tomography (PET), will allow to reliably evaluate the efficacy of therapeutic interventions in HD. This study aimed to assess the radiation burden of [F]CHDI-650, a novel fluorinated mHTT radioligand, in humans based on both in vivo and ex vivo biodistribution in mice and subsequent determination of dosimetry for dosing in humans.

View Article and Find Full Text PDF

. The International Commission on Radiological Protection (ICRP) decided to develop pregnant-female reference computational phantoms, including the maternal and fetal phantoms, through its 2007 general recommendations. Acknowledging the advantages of the mesh geometry, the ICRP decided to develop the pregnant-female mesh-type reference computational phantoms (MRCPs) for 8, 10, 15, 20, 25, 30, 35, and 38 week fetal ages directly in the mesh format.

View Article and Find Full Text PDF

Comparison of fluorescence intensity of protoporphyrin IX as observed on the screen of different cystoscopic systems.

Photodiagnosis Photodyn Ther

November 2024

Department of Urology, Kochi Medical School, Kohasu, Oko, Nankoku, Kochi, Japan; Center for Photodynamic Medicine, Kochi Medical School, Kohasu, Oko, Nankoku, Kochi, Japan.

Background: Photodynamic diagnosis-assisted transurethral resection of bladder tumor (PDD-TURBT) is an important and useful intervention used for removing non-invasive bladder cancer (NMIBC). Several PDD devices can be used in clinical practice, but few reports have compared them.In this study, we examined the differences in detected fluorescence intensity for each PDD device.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!