  • * Results showed that while the accuracy of detecting regional wall-motion abnormalities (RWMAs) improved slightly with the slow-motion technique (87.5%) compared to conventional ESE (81.0%), this difference was not statistically significant.
  • * The new method demonstrated enhanced image readability and better interreader agreement among cardiologists, suggesting it could be a beneficial tool to assist physicians in evaluations, despite the accuracy not being significantly improved.

Article Abstract

The diagnostic accuracy of exercise stress echocardiography (ESE) for myocardial ischemia requires improvement, given that it currently depends on the physicians' experience and image quality. To address this issue, we aimed to develop artificial intelligence (AI)-based slow-motion echocardiography using inter-image interpolation. The clinical usefulness of this method was evaluated for detecting regional wall-motion abnormalities (RWMAs). In this study, an AI-based echocardiographic image-interpolation pipeline was developed using optical flow calculation and prediction for in-between images. The accuracy for detecting RWMAs and image readability among 25 patients with RWMA and 25 healthy volunteers was compared between four cardiologists using slow-motion and conventional ESE. Slow-motion echocardiography was successfully developed for arbitrary time-steps (e.g., 0.125×, and 0.5×) using 1,334 videos. The RWMA detection accuracy showed a numerical improvement, but it was not statistically significant (87.5% in slow-motion echocardiography vs. 81.0% in conventional ESE; odds ratio: 1.43 [95% CI: 0.78-2.62], p = 0.25). Interreader agreement analysis (Fleiss's Kappa) for detecting RWMAs among the four cardiologists were 0.66 (95%CI: 0.55-0.77) for slow-motion ESE and 0.53 (95%CI: 0.42-0.65) for conventional ESE. Additionally, subjective evaluations of image readability using a four-point scale showed a significant improvement for slow-motion echocardiography (2.11 ± 0.73 vs. 1.70 ± 0.78, p < 0.001).In conclusion, we successfully developed slow-motion echocardiography using in-between echocardiographic image interpolation. Although the accuracy for detecting RWMAs did not show a significant improvement with this method, we observed enhanced image readability and interreader agreement. This AI-based approach holds promise in supporting physicians' evaluations.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s10554-023-02997-6DOI Listing

Publication Analysis

Top Keywords

slow-motion echocardiography
20
conventional ese
12
detecting rwmas
8
image readability
8
slow-motion
7
echocardiography
6
ese
5
development artificial
4
artificial intelligence-based
4
intelligence-based slow-motion
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!