Simulation-guided development of an optical calorimeter for high dose rate dosimetry.

Phys Eng Sci Med

School of Physical and Chemical Sciences, University of Canterbury, Christchurch, 8041, New Zealand.

Published: March 2024

AI Article Synopsis

  • Optical Calorimetry (OC) uses interferometry to measure radiation-induced changes in refractive index, allowing direct measurement of absorbed doses in water.
  • The study optimized an OC system for ultra-high dose rate applications, utilizing software for design and testing a prototype in a clinical setting.
  • Key improvements included reducing noise, isolating the system from vibrations, and controlling temperature, leading to enhanced spatial resolution and lower dose uncertainty, making OC a promising method for advanced dosimetry in high dose rate treatments.

Article Abstract

Optical Calorimetry (OC) is based on interferometry and provides a direct measurement of spatially resolved absorbed dose to water by measuring refractive index changes induced by radiation. The purpose of this work was to optimize and characterize in software an OC system tailored for ultra-high dose rate applications and to build and test a prototype in a clinical environment. A radiation dosimeter using the principles of OC was designed in optical modelling software. Traditional image quality instruments, fencepost and contrast phantoms, were utilized both in software and experimentally in a lab environment to investigate noise reduction techniques and to test the spatial and dose resolution of the system. Absolute dose uncertainty was assessed by measurements in a clinical 6 MV Flattening Filter Free (FFF) photon beam with dose rates in the range 0.2-6 Gy/s achieved via changing the distance from the source. Design improvements included: equalizing the pathlengths of the interferometer, isolating the system from external vibrations and controlling the system's internal temperature as well as application of mathematical noise reduction techniques. Simulations showed that these improvements should increase the spatial resolution from 22 to 35 lp/mm and achieve a minimum detectable dose of 0.2 Gy, which was confirmed experimentally. In the FFF beam, the absolute dose uncertainty was dose rate dependent and decreased from 2.5 ± 0.8 to 2.5 ± 0.2 Gy for dose rates of 0.2 and 6 Gy/s, respectively. A radiation dosimeter utilizing the principles of OC was developed and constructed. Optical modelling software and image quality phantoms allowed for iterative testing and refinement. The refined OC system proved capable of measuring absorbed dose to water in a linac generated photon beam. Reduced uncertainty at higher dose rates indicates the potential for OC as a dosimetry system for high dose rate techniques such as microbeam and ultra-high dose-rate radiotherapy.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s13246-023-01349-xDOI Listing

Publication Analysis

Top Keywords

dose rate
16
dose
13
dose rates
12
high dose
8
absorbed dose
8
dose water
8
radiation dosimeter
8
optical modelling
8
modelling software
8
image quality
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!