CT-based evaluation of tissue expansion in cryoablation of kidney.

Biomed Tech (Berl)

Institute of Diagnostic and Interventional Radiology, University Hospital, Goethe University Frankfurt, Frankfurt, Germany.

Published: April 2024

Objectives: To evaluate tissue expansion during cryoablation, the displacement of markers in kidney tissue was determined using computed tomographic (CT) imaging.

Methods: CT-guided cryoablation was performed in nine porcine kidneys over a 10 min period. Markers and fiber optic temperature probes were positioned perpendicular to the cryoprobe shaft in an axial orientation. The temperature measurement was performed simultaneously with the acquisitions of the CT images in 5 s intervals. The distance change of the markers to the cryoprobe was determined in each CT image and equated to the measured temperature at the marker.

Results: The greatest increase in the distance between the markers and the cryoprobe was observed in the initial phase of cryoablation. The maximum displacement of the markers was determined to be 0.31±0.2 mm and 2.8±0.02 %, respectively.

Conclusions: The mean expansion of kidney tissue during cryoablation with a single cryoprobe is 0.31±0.2 mm. The results can be used for identification of basic parameters for optimization of therapy planning.

Download full-text PDF

Source
http://dx.doi.org/10.1515/bmt-2023-0174DOI Listing

Publication Analysis

Top Keywords

tissue expansion
8
expansion cryoablation
8
displacement markers
8
kidney tissue
8
markers cryoprobe
8
cryoablation
5
markers
5
ct-based evaluation
4
tissue
4
evaluation tissue
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!