Extended duration of normothermic machine perfusion (NMP) provides opportunities to resuscitate suboptimal donor livers. This intervention requires adequate oxygen delivery typically provided by a blood-based perfusion solution. Methaemoglobin (MetHb) results from the oxidation of iron within haemoglobin and represents a serious problem in perfusions lasting > 24 h. We explored the effects of anti-oxidant, N-acetylcysteine (NAC) on the accumulation of methaemoglobin. NMP was performed on nine human donor livers declined for transplantation: three were perfused without NAC (no-NAC group), and six organs perfused with an initial NAC bolus, followed by continuous infusion (NAC group), with hourly methaemoglobin perfusate measurements. In-vitro experiments examined the impact of NAC (3 mg) on red cells (30 ml) in the absence of liver tissue. The no-NAC group sustained perfusions for an average of 96 (range 87-102) h, universally developing methaemoglobinaemia (≥ 2%) observed after an average of 45 h, with subsequent steep rise. The NAC group was perfused for an average of 148 (range 90-184) h. Only 2 livers developed methaemoglobinaemia (peak MetHb of 6%), with an average onset of 116.5 h. Addition of NAC efficiently limits formation and accumulation of methaemoglobin during NMP, and allows the significant extension of perfusion duration.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10624848PMC
http://dx.doi.org/10.1038/s41598-023-45206-zDOI Listing

Publication Analysis

Top Keywords

machine perfusion
8
donor livers
8
accumulation methaemoglobin
8
methaemoglobin nmp
8
no-nac group
8
nac group
8
nac
7
n-acetylcysteine novel
4
novel approach
4
approach methaemoglobinaemia
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!