Background: A compact PET/SPECT/CT system Inliview-3000B has been developed to provide multi-modality information on small animals for biomedical research. Its PET subsystem employed a dual-layer-offset detector design for depth-of-interaction capability and higher detection efficiency, but the irregular design caused some difficulties in calculating the normalization factors and the sensitivity map. Besides, the relatively larger (2 mm) crystal cross-section size also posed a challenge to high-resolution image reconstruction.

Purpose: We present an efficient image reconstruction method to achieve high imaging performance for the PET subsystem of Inliview-3000B.

Methods: List mode reconstruction with efficient system modeling was used for the PET imaging. We adopt an on-the-fly multi-ray tracing method with random crystal sampling to model the solid angle, crystal penetration and object attenuation effect, and modify the system response model during each iteration to improve the reconstruction performance and computational efficiency. We estimate crystal efficiency with a novel iterative approach that combines measured cylinder phantom data with simulated line-of-response (LOR)-based factors for normalization correction before reconstruction. Since it is necessary to calculate normalization factors and the sensitivity map, we stack the two crystal layers together and extend the conventional data organization method here to index all useful LORs. Simulations and experiments were performed to demonstrate the feasibility and advantage of the proposed method.

Results: Simulation results showed that the iterative algorithm for crystal efficiency estimation could achieve good accuracy. NEMA image quality phantom studies have demonstrated the superiority of random sampling, which is able to achieve good imaging performance with much less computation than traditional uniform sampling. In the spatial resolution evaluation based on the mini-Derenzo phantom, 1.1 mm hot rods could be identified with the proposed reconstruction method. Reconstruction of double mice and a rat showed good spatial resolution and a high signal-to-noise ratio, and organs with higher uptake could be recognized well.

Conclusion: The results validated the superiority of introducing randomness into reconstruction, and demonstrated its reliability for high-performance imaging. The Inliview-3000B PET subsystem with the proposed image reconstruction can provide rich and detailed information on small animals for preclinical research.

Download full-text PDF

Source
http://dx.doi.org/10.1002/mp.16814DOI Listing

Publication Analysis

Top Keywords

image reconstruction
12
pet subsystem
12
reconstruction
9
efficient image
8
dual-layer-offset detector
8
detector design
8
small animals
8
normalization factors
8
factors sensitivity
8
sensitivity map
8

Similar Publications

AAPM Truth-based CT (TrueCT) reconstruction grand challenge.

Med Phys

January 2025

Center for Virtual Imaging Trial, Carl E. Ravin Advanced Imaging Laboratories, Department of Radiology, Duke University School of Medicine, Durham, North Carolina, USA.

Background: This Special Report summarizes the 2022, AAPM grand challenge on Truth-based CT image reconstruction.

Purpose: To provide an objective framework for evaluating CT reconstruction methods using virtual imaging resources consisting of a library of simulated CT projection images of a population of human models with various diseases.

Methods: Two hundred unique anthropomorphic, computational models were created with varied diseases consisting of 67 emphysema, 67 lung lesions, and 66 liver lesions.

View Article and Find Full Text PDF

The hippocampus is a small, yet intricate seahorse-shaped tiny structure located deep within the brain's medial temporal lobe. It is a crucial component of the limbic system, which is responsible for regulating emotions, memory, and spatial navigation. This research focuses on automatic hippocampus segmentation from Magnetic Resonance (MR) images of a human head with high accuracy and fewer false positive and false negative rates.

View Article and Find Full Text PDF

Impacts of immersive 3D videos on students' surgical learning compared to 2D videos: a randomized controlled trial.

Int J Surg

December 2024

Department of Gynecological and Breast Surgery and Oncology, Pitié-Salpêtrière, Assistance Publique des Hôpitaux de Paris (AP-HP), University Hospital, Paris, France.

Background: Unlike other medical specialties, surgery is primarily learned through apprenticeship, by observing surgeons in action during operations. However, the increasing number of medical students and work-hour restrictions limit opportunities for learning in the operating room (OR). These circumstances call for novel technologies, such as immersive video.

View Article and Find Full Text PDF

Cranioventral pulmonary consolidation (CVPC) is a common lesion observed in the lungs of slaughtered pigs, often associated with Mycoplasma (M.) hyopneumoniae infection. There is a need to implement simple, fast, and valid CVPC scoring methods.

View Article and Find Full Text PDF

Background: Three-dimensional (3D) visualization has become increasingly prevalent in orthopedic education to tackle the distinct anatomical challenges of the field. However, there is a conspicuous lack of systematic reviews that thoroughly evaluate both the advantages and drawbacks of integrating 3D with problem-based learning (3D + PBL).

Methods: A rigorous search of English databases (Cochrane Library, Embase, PubMed, Scopus, and Web of Science) and Chinese databases (National Knowledge Infrastructure: CNKI, Chongqing VIP: VIP, and Wan Fang) were performed up to July 2024 to identify relevant studies.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!