Background And Purpose: Radiostereometric analysis (RSA) is the gold standard for evaluation of migration of implants. CT-RSA has been shown to have precision at the level of RSA in hip, shoulder, and knee joint replacements. We aimed to assess the impact of dose reduction on precision of CT-RSA on tibial implants, comparing it with previously published data on precision of standard dose CT-RSA on tibial implants.
Material And Methods: We performed a total knee arthroplasty on a porcine knee cadaver, and subsequent CT-RSA with low effective doses (0.02 mSv). We compared the results with previously published CT-RSA data with standard (0.08 mSv) dose. The primary outcome variable was the difference in precision of the maximum total translation (MTT). Secondary variables included ratios of variances and standard deviations, and precision of peripheral point translations, center-of-mass translations, and rotations. A difference of more than 0.1 mm in precision was defined as clinically relevant. Our hypothesis was that precisions of low and standard CT-RSA doses were equal.
Results: Low dose (mean 0.07, 95% confidence interval [CI] 0.06-0.08) and standard dose CT-RSA (0.08, CI 0.07-0.09) achieve similar precision, with difference in precision of MTT of 0.01, CI 0.00-0.02 mm. The F-statistic (0.99, CI 0.63-1.55) and sdtest (1.05, CI 0.43-2.58) also supported this.
Conclusion: We conclude that the precision of low dose CT-RSA for tibial implants on a porcine cadaver is equal to standard dose CT-RSA. However, these findings should be confirmed in clinical trials.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10618912 | PMC |
http://dx.doi.org/10.2340/17453674.2023.24022 | DOI Listing |
Acta Orthop
January 2025
Department of Orthopaedics, Institute of Clinical Sciences, Sahlgrenska Academy, University of Gothenburg, Gothenburg; Department of Orthopaedics, Sahlgrenska University Hospital, Region Västra Götaland, Gothenburg, Sweden.
Background And Purpose: Computed tomography radiostereometric analysis (CT-RSA) assesses implant micromovements using low-dose CT scans. We aimed to investigate whether CT-RSA is comparable to marker-based radiostereometric analysis (RSA) measuring early femoral head migration in cemented stems. We hypothesized that CT-RSA is comparable to marker-based RSA in evaluating femoral head subsidence.
View Article and Find Full Text PDFJ Orthop Res
December 2024
Department of Orthopedics, Institute of Clinical Sciences, Sahlgrenska Academy, University of Gothenburg, Gothenburg, Sweden.
Radiostereometric Analysis (RSA) is the most accurate method for determining early micromotions of orthopedic implants. Computed Tomography Radiostereometric Analysis (CT-RSA) is a method that can be used to determine implant and bone micromovements using low-dose CT scans. This study aimed to evaluate the reliability of the CT-RSA method in measuring the interfragmental mobility in patients who have undergone a correction osteotomy due to a malunited distal radius fracture.
View Article and Find Full Text PDFActa Orthop
January 2024
Department of Orthopedics, Skåne University Hospital, Clinical Sciences, Lund University, Lund.
Background And Purpose: Radiostereometry (RSA) is the current gold standard for evaluating early implant migration. CT-based migration analysis is a promising method, with fewer handling requirements compared with RSA and no need for implanted bone-markers. We aimed to evaluate agreement between a new artificial intelligence (AI)-based CT-RSA and model-based RSA (MBRSA) in measuring migration of cup and stem in total hip arthroplasty (THA).
View Article and Find Full Text PDFActa Orthop
October 2023
Department of Radiology and Nuclear Medicine, Oslo University Hospital Ullevål, Oslo, Norway.
Background And Purpose: Radiostereometric analysis (RSA) is the gold standard for evaluation of migration of implants. CT-RSA has been shown to have precision at the level of RSA in hip, shoulder, and knee joint replacements. We aimed to assess the impact of dose reduction on precision of CT-RSA on tibial implants, comparing it with previously published data on precision of standard dose CT-RSA on tibial implants.
View Article and Find Full Text PDFActa Orthop
July 2023
Department of Surgical Sciences, Orthopaedics, Uppsala University Hospital, Sweden.
Early implant migration is an indicator of the long-term survival/failure of implants. CT-based radio-stereometric analysis (CT-RSA) is a precise method for measuring and visualizing implant migration in vivo using image processing of CT scans. This makes the method widely applicable to orthopedic researcher.