It is very important to keep track of decreases in the bone mineral density (BMD) of elderly people since it can be correlated with the risk of incidence of major osteoporotic fractures leading to fatal injuries. Even though dual-energy X-ray absorptiometry (DXA) is the one of the most precise measuring techniques used to quantify BMD, most patients have restricted access to this machine due to high cost of DXA equipment, which is also rarely distributed to local clinics. Meanwhile, the conventional X-rays, which are commonly used for visualizing conditions and injuries due to their low cost, combine the absorption of both soft and bone tissues, consequently limiting its ability to measure BMD. Therefore, we have proposed a specialized automated smart system to quantitatively predict BMD based on a conventional X-ray image only by reducing the soft tissue effect supported by the implementation of a convolutional autoencoder, which is trained using proposed synthesized data to generate grayscale values of bone tissue alone. From the enhanced image, multiple features are calculated from the hip X-ray to predict the BMD values. The performance of the proposed method has been validated through comparison with the DXA value, which shows high consistency with correlation coefficient of 0.81 and mean absolute error of 0.069 g/cm.